EM TEST (Switzerland) GmbH
PFM 200N100.1
User’s Guide
Ver. 1.12
40 / 43
11.
Accessories for LV 124
11.1. CA LV124 Calibration Load
The calibration loads in the CA LV124 series are to be used for pulse verification of E-10 and E-13
according to LV 124: 2013-02. They consist of low-inductive resistors. The user connects the loads
directly to the outputs of the battery or data lines.
Calibration power supply
CA LV124-P1R
1
Ω
Load resistor
CA LV124-P100R
100
Ω
Load resistor
BNC cable
BNC Measuring cable 2 m
Calibration data lines
CA LV124-D1R
1
Ω
Load resistor
CA LV124-D1000R 1000
Ω
Load resistor
Supply cable
2 cables 0.5m red/black
Figure 11.1: Calibration Kit CA LV124
11.1.1. CA LV124 Calibration loads
The calibration loads of the series CA LV124 are suitable for the pulse verification as per LV 124:
2013-02 tests E-10 and E-13. They consist of four low-inductance resistors. The user connects the
calibration load directly to the power output or I/O data plugs.
Technical Data CA LV 124
Use
Battery Lines
Signal or Data Lines
Figure 11.2
Figure 11.3
Description
CA LV124-P1R
CA LV124-P100R
CA LV124-D1R
CA LV124-D1000R
Resistance
1 Ω
100 Ω
1 Ω
1.000
Ω
Accuracy
± 1 %
± 1 %
± 2 %
± 2 %
Max. Voltage
12 V
100 V
2 V
40 V
Power max.
30 W
30 W
1 W
1 W
Power peak
150 W, 1 s
100 W, 1 s
4 W, 1 s
4 W, 1 s
Dimensions
185 x 105 x 58 mm
185 x 105 x 58 mm
120 x 65 x 40 mm
120 x 65 x 40 mm
Weight
1.05 kg
0.90 kg
0.15 kg
0.15 kg