This test list can be processed with standard commercially available test aids and tools, but does not replace the inspections and
maintenance work carried out by the manufacturer.
5503.400 Oxylog 1000
Test List 2
Page 31
K5503400TL2.fm
15.03.06
Copyright reserved
. Re
prod
uc
tion perm
itted
for non-commercial pur
p
o
ses
only
.
2.6.4
Checking the Alarm Sound Silence
Test set-up:
Fig.
9:
Oxylog 1000 with test set-up: Checking the alarm sound silence
• Set the minute volume adjuster (MV) to 10 L/min.
• Set the frequency adjuster (Freq.) to 10 1/min.
• Disconnect the test lung from the ventilation valve.
Oxylog 1000 should generate an audible alarm. The "Paw low" indicator
should turn red.
• During the inspiratory phase, press the silence button.
The audible alarm should stop. The silence indicator should turn yellow.
The audible alarm should restart after 2 minutes, at the latest.
O
2
Test pressure regulator
Test lung
Ventilation valve
Содержание Oxylog 1000
Страница 2: ......
Страница 5: ...1 General ...
Страница 6: ...2 ...
Страница 9: ...5 Function Description ...
Страница 10: ...6 ...
Страница 35: ...31 Maintenance Procedures ...
Страница 36: ...32 ...
Страница 45: ...41 Schematics and Diagrams ...
Страница 46: ...42 ...
Страница 55: ...Annex Parts catalog Test List ...
Страница 56: ......
Страница 58: ...2 ...
Страница 64: ...8 ...
Страница 65: ...Test List ...
Страница 66: ......
Страница 103: ......