![Dräger Medical Oxylog 1000 Скачать руководство пользователя страница 100](http://html.mh-extra.com/html/drager-medical-0/oxylog-1000/oxylog-1000_technical-documentation-manual_2527049100.webp)
This test list can be processed with standard commercially available test aids and tools, but does not replace the inspections and
maintenance work carried out by the manufacturer.
5503.400 Oxylog 1000
Test List 2
Page 32
K5503400TL2.fm
15.03.06
Copyright reserved
. Re
prod
uc
tion perm
itted
for non-commercial pur
p
o
ses
only
.
2.7
Checking the O
2
Concentration
Test set-up:
Fig. 10:
Oxylog 1000 with test set-up: Checking the O
2
concentration
• Set the frequency adjuster (Freq.) to 10 1/min.
• Set the airway pressure (Paw) to right end stop.
• Set the Air Mix/ No Air Mix selector to the "Air Mix" position.
• Set the minute volume adjuster (MV) to 7 L/min.
The oxygen analyzer (Oxydig) should show 55 vol.% O
2
(40 to 65 vol.% O
2
)
on its display.
• Set the minute volume adjuster (MV) to 20 L/min.
The oxygen analyzer (Oxydig) should show 55 vol.% O
2
(40 to 65 vol.% O
2
)
on its display.
• Set the Air Mix/ No Air Mix selector to the "No Air Mix" position.
The oxygen analyzer (Oxydig) should show 100 vol.% O
2
on its display.
• Switch off Oxylog 1000 using the ON/OFF switch.
+
Vol%
O2
100
21
Cal
T/
l
O
2
Test pressure regulator
Test lung
Ventilation valve
Test adapter
O
2
analyzer
Содержание Oxylog 1000
Страница 2: ......
Страница 5: ...1 General ...
Страница 6: ...2 ...
Страница 9: ...5 Function Description ...
Страница 10: ...6 ...
Страница 35: ...31 Maintenance Procedures ...
Страница 36: ...32 ...
Страница 45: ...41 Schematics and Diagrams ...
Страница 46: ...42 ...
Страница 55: ...Annex Parts catalog Test List ...
Страница 56: ......
Страница 58: ...2 ...
Страница 64: ...8 ...
Страница 65: ...Test List ...
Страница 66: ......
Страница 103: ......