This test list can be processed with standard commercially available test aids and tools, but does not replace the inspections and
maintenance work carried out by the manufacturer.
5503.400 Oxylog 1000
Test List
Page 9
K5503400TL1.fm
15.03.06
Co
pyri
ght
reser
ve
d
.
Repro
ducti
on p
e
rmitt
e
d o
n
ly f
o
r non-commerc
ial
purpo
ses.
2.2
Testing vent valve
Test set-up:
Fig.
4:
Oxylog 1000 with test set-up: Testing the vent valve
• With the test pressure reducer feed a flow of 10 L/min into the
connection socket for the ventilation hose.
The maximum pressure rise is 8 mbar (hPa) (read off the pressure rise from
the pressure gauge of the Oxylog 1000).
CAUTION
To check the vent valve, do not connect the compressed gas supply to the
connection socket for compressed gas supply and switch off the
Oxylog 1000 at the main switch!
O2
Test pressure reducer
Flowmeter 0 to
10 L/min
Содержание Oxylog 1000
Страница 2: ......
Страница 5: ...1 General ...
Страница 6: ...2 ...
Страница 9: ...5 Function Description ...
Страница 10: ...6 ...
Страница 35: ...31 Maintenance Procedures ...
Страница 36: ...32 ...
Страница 45: ...41 Schematics and Diagrams ...
Страница 46: ...42 ...
Страница 55: ...Annex Parts catalog Test List ...
Страница 56: ......
Страница 58: ...2 ...
Страница 64: ...8 ...
Страница 65: ...Test List ...
Страница 66: ......
Страница 103: ......