Function Description
3.5
Alarm Silence
When the pushbutton valve (T1) is pressed the 3/2-way valve (DMR15) fills
the volume (V3) and activates the 3/2-way valve (DMR16). The 3/2-way valve
(DMR17) vents the alarm generator to atmosphere. The alarm generator (L1)
is inactive up to the switching point of the volume (V3) (approx. 2 minutes).
Fig.
9:
Segment of the flow chart: Alarm silence
DMR15
DMR16
DMR17
V3
L1
Alarm silence
T1