This test list can be processed with standard commercially available test aids and tools, but does not replace the inspections and
maintenance work carried out by the manufacturer.
5503.400 Oxylog 1000
Test List 2
Page 28
K5503400TL2.fm
15.03.06
Copyright reserved
. Re
prod
uc
tion perm
itted
for non-commercial pur
p
o
ses
only
.
2.6
Checking Warning Logics and Silence Control
2.6.1
Checking the Supply Pressure Warning (Pv)
Test set-up:
Fig.
6:
Oxylog 1000 with test set-up for supply pressure warning check
• Set the minute volume adjuster (MV) to left limit stop.
• Set the frequency adjuster (Freq.) to 10 1/min.
• Set the airway pressure adjuster (Paw) to approx. 40 mbar (hPa).
• Set the Air Mix/ No Air Mix selector to the "No Air Mix" position.
• Slowly close the pressure regulator of the test pressure reducer.
• Watch the pressure gauge of the test pressure reducer.
As soon as the downstream pressure reaches 2.2 bar to 1.8 bar, the "Pv"
indicator turns red. The supply pressure is insufficient.
• Slowly open the pressure regulator of the test pressure reducer.
As soon as the downstream pressure reaches 2.3 bar to 2.7 bar, the "Pv"
indicator turns green. The supply pressure is sufficient.
O
2
Test pressure regulator
Test lung
Ventilation valve
Содержание Oxylog 1000
Страница 2: ......
Страница 5: ...1 General ...
Страница 6: ...2 ...
Страница 9: ...5 Function Description ...
Страница 10: ...6 ...
Страница 35: ...31 Maintenance Procedures ...
Страница 36: ...32 ...
Страница 45: ...41 Schematics and Diagrams ...
Страница 46: ...42 ...
Страница 55: ...Annex Parts catalog Test List ...
Страница 56: ......
Страница 58: ...2 ...
Страница 64: ...8 ...
Страница 65: ...Test List ...
Страница 66: ......
Страница 103: ......