![Dräger Medical Oxylog 1000 Скачать руководство пользователя страница 84](http://html.mh-extra.com/html/drager-medical-0/oxylog-1000/oxylog-1000_technical-documentation-manual_2527049084.webp)
This test list can be processed with standard commercially available test aids and tools, but does not replace the inspections and
maintenance work carried out by the manufacturer.
5503.400 Oxylog 1000
Test List
Page 16
K5503400TL1.fm
15.03.06
Co
pyri
ght
reser
ve
d
.
Repro
ducti
on p
e
rmitt
e
d o
n
ly f
o
r non-commerc
ial
purpo
ses.
2.7
Testing O
2
concentration
Test set-up:
Fig. 10:
Oxylog 1000 with test set-up: Testing the O
2
concentration
• Set the frequency selector (Freq.) to 10 1/min.
• Set the ventilation pressure (Paw) to full right.
• Set the Air Mix/No Air Mix selector switch to „Air Mix“.
• Set the minute volume adjuster (MV) to 7 L/min.
The Oxydig shows 55 Vol.%O
2
(± 17 Vol.%O
2
) on the display.
• Set the minute volume adjuster (MV) to 20 L/min.
The Oxydig shows 55 Vol.%O
2
(± 17 Vol.%O
2
) on the display.
• Set the Air Mix/No Air Mix selector switch to „No Air Mix“.
The Oxydig shows 100 Vol.%O
2
on the display.
• Switch off the Oxylog 1000 at the main switch.
+
Vol%
O2
100
21
Cal
T/
l
O2
Test pressure
reducer
Test lung
Ventilation valve
Test adapter
O2 analyser
Содержание Oxylog 1000
Страница 2: ......
Страница 5: ...1 General ...
Страница 6: ...2 ...
Страница 9: ...5 Function Description ...
Страница 10: ...6 ...
Страница 35: ...31 Maintenance Procedures ...
Страница 36: ...32 ...
Страница 45: ...41 Schematics and Diagrams ...
Страница 46: ...42 ...
Страница 55: ...Annex Parts catalog Test List ...
Страница 56: ......
Страница 58: ...2 ...
Страница 64: ...8 ...
Страница 65: ...Test List ...
Страница 66: ......
Страница 103: ......