This test list can be processed with standard commercially available test aids and tools, but does not replace the inspections and
maintenance work carried out by the manufacturer.
5503.400 Oxylog 1000
Test List 2
Page 27
K5503400TL2.fm
15.03.06
Copyright reserved
. Re
prod
uc
tion perm
itted
for non-commercial pur
p
o
ses
only
.
2.5
Checking the Minute Volume
Test set-up:
Fig.
5:
Oxylog 1000 with test set-up: Checking the minute volume
• Set the airway pressure adjuster (Paw) to right end stop.
• Set the frequency adjuster (Freq.) to 10 1/min.
• Set the minute volume adjuster (MV) to various steps, see the following
table.
• Set the required "counterpressure" during the inspiratory phase using
the adjustable restrictor. (Read counterpressure on pressure gauge of
Oxylog 1000).
• Measure the minute volume with and without ambient-air enrichment,
see the following table.
Minute volume
adjuster
Flow
Counterpressure
setting
(mbar/hPa)
Selector in
No Air Mix position
Selector in
Air Mix
position
Minimum
7,5
15 to 40
6 to 9 L/min
—
Minimum
8
no
counterpressure
— 6
to
10 L/min
Maximum
50
15 to 40
40 to 60 L/min
—
Maximum
50
no
counterpressure
— 40
to
60 L/min
O
2
Test pressure regulator
Flowmeter
Adjustable
restrictor
Содержание Oxylog 1000
Страница 2: ......
Страница 5: ...1 General ...
Страница 6: ...2 ...
Страница 9: ...5 Function Description ...
Страница 10: ...6 ...
Страница 35: ...31 Maintenance Procedures ...
Страница 36: ...32 ...
Страница 45: ...41 Schematics and Diagrams ...
Страница 46: ...42 ...
Страница 55: ...Annex Parts catalog Test List ...
Страница 56: ......
Страница 58: ...2 ...
Страница 64: ...8 ...
Страница 65: ...Test List ...
Страница 66: ......
Страница 103: ......