Function Description
Oxylog 1000
3.3
Air Mix/No Air Mix
Selector Switch
The setting of the Air Mix/No Air Mix selector switch determines the O
2
con-
centration of the supply gas for the patient.
When the Air Mix/No Air Mix selector switch is set to "Air Mix", the injector (I)
takes in ambient air through the 3/2-way valve (WV2). The O
2
concentration
of the supply gas for the patient is approx. 50 vol.% O
2
.
When the Air Mix/No Air Mix selector switch is set to "No Air Mix", the injector
(I) takes in the supply gas through the 3/2-way valve (WV2). The O
2
concen-
tration of the supply gas for the patient is approx. 100 vol.% O
2
.
Figure 5
Segment of the flow chart: Air Mix/No Air Mix selector switch
WV2
S2
I
Содержание Oxylog 1000
Страница 2: ......
Страница 5: ...1 General ...
Страница 6: ...2 ...
Страница 9: ...5 Function Description ...
Страница 10: ...6 ...
Страница 35: ...31 Maintenance Procedures ...
Страница 36: ...32 ...
Страница 45: ...41 Schematics and Diagrams ...
Страница 46: ...42 ...
Страница 55: ...Annex Parts catalog Test List ...
Страница 56: ......
Страница 58: ...2 ...
Страница 64: ...8 ...
Страница 65: ...Test List ...
Страница 66: ......
Страница 103: ......