This test list can be processed with standard commercially available test aids and tools, but does not replace the inspections and
maintenance work carried out by the manufacturer.
5503.400 Oxylog 1000
Test List
Page 13
K5503400TL1.fm
15.03.06
Co
pyri
ght
reser
ve
d
.
Repro
ducti
on p
e
rmitt
e
d o
n
ly f
o
r non-commerc
ial
purpo
ses.
2.6.2
Testing disconnect alarm
Test set-up:
Fig.
7:
Oxylog 1000 with test set-up: Testing the disconnect alarm
• Set the minute volume adjuster (MV) to 10 L/min.
• Set the frequency selector (Freq.) to 10 1/min.
• Set the Air Mix/No Air Mix selector switch to „Air Mix“.
• Set the ventilation pressure selector (Paw) to full right.
• During the inhalation phase set a ventilation pressure of approx.
30 mbar (hPa) with the adjustable resistor. (Note: Read off the
ventilation pressure from the pressure gauge of the Oxylog 1000).
• Slowly open the adjustable resistor during the inhalation phase until the
„rot“ indicator (Paw low) lights up.
The „Paw low“ switching point is at approx. 10 mbar (hPa). (Read off the
switching point from the pressure gauge of the Oxylog 1000 and from the
separate pressure gauge).
O2
Test pressure
reducer
Ventilation hose
Adjustable
resistor
Separate
pressure gauge
T-piece
Содержание Oxylog 1000
Страница 2: ......
Страница 5: ...1 General ...
Страница 6: ...2 ...
Страница 9: ...5 Function Description ...
Страница 10: ...6 ...
Страница 35: ...31 Maintenance Procedures ...
Страница 36: ...32 ...
Страница 45: ...41 Schematics and Diagrams ...
Страница 46: ...42 ...
Страница 55: ...Annex Parts catalog Test List ...
Страница 56: ......
Страница 58: ...2 ...
Страница 64: ...8 ...
Страница 65: ...Test List ...
Страница 66: ......
Страница 103: ......