This test list can be processed with standard commercially available test aids and tools, but does not replace the inspections and
maintenance work carried out by the manufacturer.
5503.400 Oxylog 1000
Test List
Page 11
K5503400TL1.fm
15.03.06
Co
pyri
ght
reser
ve
d
.
Repro
ducti
on p
e
rmitt
e
d o
n
ly f
o
r non-commerc
ial
purpo
ses.
2.5
Testing minute volume
Test set-up:
Fig.
5:
Oxylog 1000 with test set-up: Testing the minute volume
• Set the ventilation pressure selector (Paw) to full right.
• Set the frequency selector (Freq.) to 10 1/min.
• Set the minute volume adjuster (MV) variably; see following table.
• Set the required „compensation pressure“ with the adjustable resistor
during the inhalation phase. (Read off the compensation pressure from
the pressure gauge of the Oxylog 1000).
• Determine the minute volume with and without enrichment of the
ambient air; see following table.
Minute volume
adjuster
Flow
Compensati
on pressure
setting
(mbar/hPa)
Selector switch set
to
No Air Mix
Selector switch
set to
Air Mix
3 L/min
7.5
approx. 2
7.5 (
±
1 L/min)
7.5 (
±
0,6L/min)
20 L/min
50
approx. 20
49 (
±
6.6 L/min)
49 (
±
7 L/min)
O2
Test pressure
reducer
Flowmeter
Adjustable
resistor
Содержание Oxylog 1000
Страница 2: ......
Страница 5: ...1 General ...
Страница 6: ...2 ...
Страница 9: ...5 Function Description ...
Страница 10: ...6 ...
Страница 35: ...31 Maintenance Procedures ...
Страница 36: ...32 ...
Страница 45: ...41 Schematics and Diagrams ...
Страница 46: ...42 ...
Страница 55: ...Annex Parts catalog Test List ...
Страница 56: ......
Страница 58: ...2 ...
Страница 64: ...8 ...
Страница 65: ...Test List ...
Страница 66: ......
Страница 103: ......