This test list can be processed with standard commercially available test aids and tools, but does not replace the inspections and
maintenance work carried out by the manufacturer.
5503.400 Oxylog 1000
Test List 2
Page 23
K5503400TL2.fm
15.03.06
Copyright reserved
. Re
prod
uc
tion perm
itted
for non-commercial pur
p
o
ses
only
.
• Connect the compressed-gas connector to the socket of the central gas
supply system/compressed-gas cylinder.
• Connect a pressure gauge to the right-hand hose of the "Pv" indicator.
• Connect ventilation hose with flowmeter (measuring range: 10 to
120 L/min) to connecting piece (ventilation hose).
Fig.
2:
Oxylog 1000 with test set-up: Checking the supply pressure
• Set the frequency adjuster (Freq.) to 10 1/min.
• Set the minute volume adjuster (MV) to right limit stop.
• Set the airway pressure (Paw) to right end stop.
• Set the Air Mix/ No Air Mix selector to the "No Air Mix" position.
• Place the ON/OFF switch of Oxylog 1000 to position "I".
At a flow of approx. 55 L/min (indicated on the flowmeter) the connected
pressure gauge should indicate a pressure of approx. 1.5 bar.
• Place the ON/OFF switch of Oxylog 1000 to position "O".
• Remove the pressure gauge.
Flowmeter 10 to
120 L/min
"Pv" indicator
ON/OFF switch
Содержание Oxylog 1000
Страница 2: ......
Страница 5: ...1 General ...
Страница 6: ...2 ...
Страница 9: ...5 Function Description ...
Страница 10: ...6 ...
Страница 35: ...31 Maintenance Procedures ...
Страница 36: ...32 ...
Страница 45: ...41 Schematics and Diagrams ...
Страница 46: ...42 ...
Страница 55: ...Annex Parts catalog Test List ...
Страница 56: ......
Страница 58: ...2 ...
Страница 64: ...8 ...
Страница 65: ...Test List ...
Страница 66: ......
Страница 103: ......