![Dräger Medical Oxylog 1000 Скачать руководство пользователя страница 88](http://html.mh-extra.com/html/drager-medical-0/oxylog-1000/oxylog-1000_technical-documentation-manual_2527049088.webp)
This test list can be processed with standard commercially available test aids and tools, but does not replace the inspections and
maintenance work carried out by the manufacturer.
5503.400 Oxylog 1000
Test List 2
Page 20
K5503400TL2.fm
15.03.06
Copyright reserved
. Re
prod
uc
tion perm
itted
for non-commercial pur
p
o
ses
only
.
Required Test Equipment
Test equipment
Range
Accuracy class
Pressure gauge
–30 to +120 mbar
(hPa)
1.6
Pressure gauge
0 to 6 bar
1.6
Flowmeter
10 to 120 L/min
1.6
Flowmeter
2 to 10 L/min
1.6
Test pressure regulator
Hoses, various
T-piece
Test lung
Oxygen analyzer, e.g. Oxydig
Adjustable restrictor
Oxygen measurement test
adapter
Stopwatch
Hose clamp
Sealing plug