This test list can be processed with standard commercially available test aids and tools, but does not replace the inspections and
maintenance work carried out by the manufacturer.
5503.400 Oxylog 1000
Test List
Page 15
K5503400TL1.fm
15.03.06
Co
pyri
ght
reser
ve
d
.
Repro
ducti
on p
e
rmitt
e
d o
n
ly f
o
r non-commerc
ial
purpo
ses.
2.6.4
Testing alarm tone muting (Silence)
Test set-up:
Fig.
9:
Oxylog 1000 with test set-up: Testing the muting (Silence)
• Set the minute volume adjuster (MV) to 10 L/min.
• Set the frequency selector (Freq.) to 10 1/min.
• Detach the test lung from the ventilation valve.
The Oxylog 1000 emits an audible alarm. The „Paw low“ indicator lights up
red.
• During the inhalation phase press the „Alarm silence“ key.
The audible alarm stops. The „Alarm silence“ indicator lights up yellow.
After a maximum of 2 minutes the audible alarm sounds again.
O2
Test pressure
reducer
Test lung
Ventilation valve
Содержание Oxylog 1000
Страница 2: ......
Страница 5: ...1 General ...
Страница 6: ...2 ...
Страница 9: ...5 Function Description ...
Страница 10: ...6 ...
Страница 35: ...31 Maintenance Procedures ...
Страница 36: ...32 ...
Страница 45: ...41 Schematics and Diagrams ...
Страница 46: ...42 ...
Страница 55: ...Annex Parts catalog Test List ...
Страница 56: ......
Страница 58: ...2 ...
Страница 64: ...8 ...
Страница 65: ...Test List ...
Страница 66: ......
Страница 103: ......