![Dräger Medical Oxylog 1000 Скачать руководство пользователя страница 43](http://html.mh-extra.com/html/drager-medical-0/oxylog-1000/oxylog-1000_technical-documentation-manual_2527049043.webp)
5503.400
F
o
r internal use only
. Copyright reserved
.
V
e
rs
io
n
2
.0_
R
e
le
a
s
ed
_Pr
in
ted
on
_15
.03
.0
6_K
55
034
00
V01
.fm
39
Oxylog 1000
Maintenance Procedures
If the measured pressure is not 1.5 bar, adjust the pressure regulator as
follows:
•
Release the adjusting ring of the pressure reducer by pulling the ring out.
•
Turn the adjusting ring clockwise to increase the downstream pressure of
the pressure regulator, or counter-clockwise to reduce the pressure.
•
After setting the correct downstream pressure, press the adjusting ring of
the pressure regulator down to prevent the downstream pressure being
unintentionally altered.
•
Set the main switch of the Oxylog 1000 to „O“.
•
Detach the hose from the pressure gauge and re-connect it to the con-
nection socket of the main switch.
•
Check that there is no leakage in the hose system of the Oxylog 1000.
•
Install the drawer unit in the reverse sequence into the housing of the
Oxylog 1000.
Содержание Oxylog 1000
Страница 2: ......
Страница 5: ...1 General ...
Страница 6: ...2 ...
Страница 9: ...5 Function Description ...
Страница 10: ...6 ...
Страница 35: ...31 Maintenance Procedures ...
Страница 36: ...32 ...
Страница 45: ...41 Schematics and Diagrams ...
Страница 46: ...42 ...
Страница 55: ...Annex Parts catalog Test List ...
Страница 56: ......
Страница 58: ...2 ...
Страница 64: ...8 ...
Страница 65: ...Test List ...
Страница 66: ......
Страница 103: ......