Function Description
Oxylog 1000
5
Pneumatic Control
(new Oxylog 1000)
5.1
Gas Supply
The compressed gas is supplied from the central supply system (CS) or from
an O
2
cylinder with pressure regulator. The operating pressure of the
Oxylog 1000 is between 2.7 bar and 6 bar.
The supply gas flows via the port (A1), the filter (F1) and the pressure regula-
tor (DR1) to the manifold (VL). The pressure regulator (DR1) regulates the
pressure of the supply gas to 1.5 bar.
Fig. 12:
Segment of the flow chart: Supply gas inlet
CAUTION
Risk of damage to equipment! Dust, oil and humidity in medical gases may
impair the functioning of the equipment or damage it. Use only dust-free,
oil-free and dry medical gases.
FD
RV1
I
WV1
WV2
MV
D3
V1
A1
F1
S1
DR1
VL
VL1
S2
Содержание Oxylog 1000
Страница 2: ......
Страница 5: ...1 General ...
Страница 6: ...2 ...
Страница 9: ...5 Function Description ...
Страница 10: ...6 ...
Страница 35: ...31 Maintenance Procedures ...
Страница 36: ...32 ...
Страница 45: ...41 Schematics and Diagrams ...
Страница 46: ...42 ...
Страница 55: ...Annex Parts catalog Test List ...
Страница 56: ......
Страница 58: ...2 ...
Страница 64: ...8 ...
Страница 65: ...Test List ...
Страница 66: ......
Страница 103: ......