Function Description
Oxylog 1000
5.4.6
Gas Supply Failure
Warning (Pv)
If the gas supply fails or falls below a defined pressure, the Oxylog 1000
delivers two alarms:
–
Visual alarm indicator "Pv"
–
Audible alarm (alarm generator)
The alarm is triggered under the following conditions:
–
The pressure of the gas supply from the central supply system or from the
O
2
cylinder is inadequate or absent
–
The valve of the O
2
cylinder is closed
–
The gas supply connector is incorrectly installed
Figure 17
Segment of the flow chart: Gas supply failure alarm
FD
V3
RV1
I
WV1
WV2
EV
SV
MV
D3
D4
RV2
V1
A1
F1
S1
DR1
VL
VL1
S2
DM
DMR18
NV
WV3
PA
BV
S
RV3
RV5
RV4
DMR7
DMR8
DMR9
DMR10
DMR12
DMR13
DMR11
DMR14
F
B
G
D1
C
M
L
N
H
P
K
DMR1
DMR3
DMR2
Содержание Oxylog 1000
Страница 2: ......
Страница 5: ...1 General ...
Страница 6: ...2 ...
Страница 9: ...5 Function Description ...
Страница 10: ...6 ...
Страница 35: ...31 Maintenance Procedures ...
Страница 36: ...32 ...
Страница 45: ...41 Schematics and Diagrams ...
Страница 46: ...42 ...
Страница 55: ...Annex Parts catalog Test List ...
Страница 56: ......
Страница 58: ...2 ...
Страница 64: ...8 ...
Страница 65: ...Test List ...
Страница 66: ......
Страница 103: ......