Function Description
Oxylog 1000
3.4.4
Disconnect Alarm
The disconnect alarm (Paw low) warns of a pressure drop in the patient gas.
The disconnect alarm is triggered under the following conditions:
–
The hose system connection to the patient is interrupted
–
The ventilation valve is incorrectly fitted or damaged
–
The cuff has a leak
–
Internal device error
If the ventilation pressure is less than 10 mbar (disconnect condition), the
3/2-way valve (DMR7 - see illustration above) is not triggered. That is, the
3/2-way valves (DMR9 and DMR10) operate the 3/2-way valve (DMR11).
During the expiration phase the indicator (Paw low) visually indicates the dis-
connection and the alarm generator (L1) delivers an audible warning.
If the ventilation pressure is higher than 10 mbar, the 3/2-way valve (DMR7)
switches. The 3/2-way valves (DMR9, DMR10 and DMR11) vent to atmo-
sphere. The visual and audible alarms are suppressed.
Figure 6
Flow chart of the Oxylog 1000: Disconnect Alarm
DMR7
DMR9
DMR10
DMR11
Ventilation pressure
L1
Содержание Oxylog 1000
Страница 2: ......
Страница 5: ...1 General ...
Страница 6: ...2 ...
Страница 9: ...5 Function Description ...
Страница 10: ...6 ...
Страница 35: ...31 Maintenance Procedures ...
Страница 36: ...32 ...
Страница 45: ...41 Schematics and Diagrams ...
Страница 46: ...42 ...
Страница 55: ...Annex Parts catalog Test List ...
Страница 56: ......
Страница 58: ...2 ...
Страница 64: ...8 ...
Страница 65: ...Test List ...
Страница 66: ......
Страница 103: ......