Schematics and Diagrams

Oxylog 1000

Table 1

Key to flow chart (old)

A1

Compressed-gas connection 

BV

Ventilation valve

D1 to D7

Restrictor

DM

Pressure gauge

DMR1 to DMR18

3/2-way valve 

DR

Pressure regulator

EV

Vent valve

F1

Filter

FD

Frequency control valve

I

Injector

AG

Alarm generator

MV

Minute volume control valve

NV

Emergency air valve

PA

Patient connection

RV1 to RV5

Non-return valve

S1

ON/OFF switch

S2

Air Mix/No Air Mix Selector

S

Hose system 

SV

Safety valve

T1

Pushbutton valve

V1 to V3

Volume

VL

Manifold

WV1

2/2-way valve (changeover valve)

WV2

3/2-way valve 

WV3

Supply pressure switching valve

Z1 to Z4

Indicator

Содержание Oxylog 1000

Страница 1: ...5503 400 9036037 Revision 4 0 Because you care Technical Documentation Oxylog 1000 Emergency Ventilator ...

Страница 2: ......

Страница 3: ...d Oxylog 1000 10 3 1 Gas Supply 10 3 2 Frequency Control 11 3 3 Air Mix No Air Mix Selector Switch 12 3 4 Safety Devices 13 3 5 Alarm Silence 17 3 6 Pressure Sensor 18 4 Location of Oxylog 1000 old Assemblies 19 5 Pneumatic Control new Oxylog 1000 20 5 1 Gas Supply 20 5 2 Frequency Control 21 5 3 Air Mix No Air Mix Selector Switch 22 5 4 Safety Devices 23 5 5 Alarm Silence 27 5 6 Pressure Sensor 2...

Страница 4: ... 2 2 Measures every six years 33 3 Pressure regulator 34 3 1 General points concerning the pressure regulator 34 3 2 Removing replacing the pressure regulator 34 Schematics and Diagrams 1 Flow chart of the Oxylog 1000 old 43 2 Flow chart of the Oxylog 1000 new 45 3 Tubing Diagram of the Oxylog 1000 old 47 4 Tubing Diagram of the Oxylog 1000 new 49 Annex Parts catalog Test List ...

Страница 5: ...1 General ...

Страница 6: ...2 ...

Страница 7: ... and may not be disclosed to third parties without the prior written consent of the manufac turer This Technical Documentation is for the purpose of information only Product descriptions found in this Technical Documentation are in no way a substitute for reading and studying the Instructions for Use Operating Manual enclosed with the product at the time of delivery WARNING A WARNING statement pro...

Страница 8: ...ough research and development and Dräger Medical reserves the right to make changes to this Technical Documentation without notice NOTE Unless otherwise stated reference is made to laws regulations or stan dards as amended applicable in the Federal Republic of Germany for equipment used or serviced in Germany Users or technicians in all other countries must verify compliance with local laws or app...

Страница 9: ...5 Function Description ...

Страница 10: ...6 ...

Страница 11: ...on Description 1 General information on the Oxylog 1000 The Oxylog 1000 is a time controlled volume constant emergency and trans port ventilator for patients weighing 7 5 kg or above The Oxylog 1000 has the operating mode controlled ventilation IPPV Intermittent Positive Pres sure Ventilation Fig 1 Front view of the Oxylog 1000 ...

Страница 12: ...of the Oxylog 1000 The Oxylog 1000 comprises the following components Pressure reducer ON OFF switch Inspiratory expiratory switching valve Minute volume control valve Frequency control Safety valve Emergency air valve Pressure gauge Injector Vent valve Air Mix No Air Mix selector switch Logic circuit for disconnect alarm and stenosis alarm Alarm silence ...

Страница 13: ...tenosis alarm Alarm generator Alarm silence Safety valve Injector Insp exp switching valve ON OFF switch Pressure regulator Frequency control Filter Air Mix No Air Mix selector switch Emergen cy air valve Pressure gauge Ventilation valve Paw low Paw high Silence P supply pressure Switching valve Supply pressure switching valve Vent valve Pressure sensor 2 7 bar 6 bar Flow control Minute volume val...

Страница 14: ...tor The operating pressure of the Oxylog 1000 is between 2 7 bar and 6 bar The supply gas flows via the port A1 the filter F1 and the pressure regula tor DR1 to the manifold O The pressure regulator DR1 regulates the pressure of the supply gas to 1 5 bar Fig 3 Segment of the flow chart Supply gas inlet CAUTION Risk of damage to equipment Dust oil and humidity in medical gases may impair the functi...

Страница 15: ...ed with a delay this vents DMR6 4 and 1 At the same time DMR3 is switched over via D1 and DMR2 and WV1 switches to inspiration Vent valve EV is closed via DMR18 Figure 4 Segment of the flow chart Frequency Control During the inspiration phase the supply gas flows through the switching valve WV1 see illustration above the minute volume control valve MV the injector I the patient connection port PA ...

Страница 16: ...tient When the Air Mix No Air Mix selector switch is set to Air Mix the injector I takes in ambient air through the 3 2 way valve WV2 The O2 concentration of the supply gas for the patient is approx 50 vol O2 When the Air Mix No Air Mix selector switch is set to No Air Mix the injector I takes in the supply gas through the 3 2 way valve WV2 The O2 concen tration of the supply gas for the patient i...

Страница 17: ... alarm Alarm warning of failure of the compressed gas supply Alarm silence 3 4 1 Mechanical Safety Valve The mechanical safety valve limits the ventilation pressure to approx 80 mbar 3 4 2 Pressure Gauge The pressure gauge indicates the current airway pressure during the inspira tion phase The display range is 10 mbar to 80 mbar 3 4 3 Emergency Air Valve The spontaneously breathing patient can bre...

Страница 18: ...ice error If the ventilation pressure is less than 10 mbar disconnect condition the 3 2 way valve DMR7 see illustration above is not triggered That is the 3 2 way valves DMR9 and DMR10 operate the 3 2 way valve DMR11 During the expiration phase the indicator Paw low visually indicates the dis connection and the alarm generator L1 delivers an audible warning If the ventilation pressure is higher th...

Страница 19: ...bar but the alarm limit 30 mbar to 60 mbar is pre set lower the 3 2 way valve DMR8 is not triggered The 3 2 way valves DMR12 and DMR13 vent through the 3 2 way valve DMR8 If the ventilation pressure exceeds the pre set alarm limit the 3 2 way valve DMR8 activates the 3 2 way valves DMR12 and DMR13 The 3 2 way valve DMR13 activates the Paw high indicator Z3 and the 3 2 way valve DMR14 activates the...

Страница 20: ...efined pressure the Oxylog 1000 delivers two alarms Visual alarm indicator Pv Audible alarm alarm generator The alarm is triggered under the following conditions The pressure of the gas supply from the central supply system or from the O2 cylinder is inadequate or absent The valve of the O2 cylinder is closed The gas supply connector is incorrectly installed Figure 8 Segment of the flow chart Gas ...

Страница 21: ...en the pushbutton valve T1 is pressed the 3 2 way valve DMR15 fills the volume V3 and activates the 3 2 way valve DMR16 The 3 2 way valve DMR17 vents the alarm generator to atmosphere The alarm generator L1 is inactive up to the switching point of the volume V3 approx 2 minutes Fig 9 Segment of the flow chart Alarm silence DMR15 DMR16 DMR17 V3 L1 Alarm silence T1 ...

Страница 22: ...dersides of the diaphragms The supply pressure Insp is cen tered on the craters on the pressure sensor during the inspiration phase The built in mechanical springs in the pressure sensor exert different forces on the diaphragms Non adjustable spring force 10 mbar Spring force adjustable 30 mbar to 60 mbar with the alarm control valve One or other of the diaphragms is raised according to the respec...

Страница 23: ... valve DMR8 10 Indicator Silence 18 Air Mix No Air Mix 3 Pressure regulator 11 Alarm control valve 19 On switch 4 Alarm logic under frequency control 12 Silence button 20 Patient block 5 Volume expiration 13 Indicator Paw low 21 Vent valve 6 Volume inspiration 14 Frequency control valve 22 DMR18 7 Alarm silence 15 Indicator Paw high 23 Patient connection port 8 Pressure sensor double diaphragm swi...

Страница 24: ...g pressure of the Oxylog 1000 is between 2 7 bar and 6 bar The supply gas flows via the port A1 the filter F1 and the pressure regula tor DR1 to the manifold VL The pressure regulator DR1 regulates the pressure of the supply gas to 1 5 bar Fig 12 Segment of the flow chart Supply gas inlet CAUTION Risk of damage to equipment Dust oil and humidity in medical gases may impair the functioning of the e...

Страница 25: ...on how the frequency adjusting valve FD is set the volume V1 is filled faster or slower thus determining the duration of the inspiration As soon as volume V1 is filled DMR1 switches thereby switching also DMR2 The gas present at DMR3 and WV1 can escape through DMR2 DMR3 and WV1 switch Expira tion starts The gas contained in the volume escapes through the frequency adjusting valve FD and DMR1 Depen...

Страница 26: ...ugh the 3 2 way valve WV2 The O2 concentration of the supply gas for the patient is approx 50 vol O2 When the Air Mix No Air Mix selector switch is set to No Air Mix the injector I takes in the supply gas through the 3 2 way valve WV2 The O2 concen tration of the supply gas for the patient is approx 100 vol O2 Figure 14 Segment of the flow chart Air Mix No Air Mix selector switch FD V3 RV1 I WV1 W...

Страница 27: ... alarm Stenosis alarm Gas supply failure alarm Alarm silence 5 4 1 Mechanical Safety Valve The mechanical safety valve limits the ventilation pressure to approx 90 mbar 5 4 2 Pressure Gauge The pressure gauge indicates the current airway pressure during the inspira tion phase The display range is 10 mbar to 80 mbar 5 4 3 Emergency Air Valve The spontaneously breathing patient can breathe in ambien...

Страница 28: ...R7 see illustration above is not triggered That is the 3 2 way valves DMR9 and DMR10 operate the 3 2 way valve DMR11 During the expiration phase the indicator Paw low visually indicates the dis connection and the alarm generator L1 delivers an audible warning If the ventilation pressure is higher than 10 mbar the 3 2 way valve DMR7 switches The 3 2 way valves DMR9 DMR10 and DMR11 vent to atmo sphe...

Страница 29: ...gered The 3 2 way valves DMR12 and DMR13 vent through the 3 2 way valve DMR8 If the ventilation pressure exceeds the pre set alarm limit the 3 2 way valve DMR8 activates the 3 2 way valves DMR12 and DMR13 The 3 2 way valve DMR13 activates the Paw high indicator Z3 and the 3 2 way valve DMR14 activates the alarm generator L1 At the same time the 3 2 way valve DMR8 switches the 3 2 way valve DMR18 T...

Страница 30: ... alarm generator The alarm is triggered under the following conditions The pressure of the gas supply from the central supply system or from the O2 cylinder is inadequate or absent The valve of the O2 cylinder is closed The gas supply connector is incorrectly installed Figure 17 Segment of the flow chart Gas supply failure alarm FD V3 RV1 I WV1 WV2 EV SV MV D3 D4 RV2 V1 A1 F1 S1 DR1 VL VL1 S2 DM D...

Страница 31: ... pushbutton valve T1 is pressed the 3 2 way valve DMR15 fills the volume V3 and activates the 3 2 way valve DMR16 The 3 2 way valve DMR17 vents the alarm generator to atmosphere The alarm generator L1 is inactive up to the switching point of the volume V3 approx 2 minutes Fig 18 Segment of the flow chart Alarm silence V3 T1 L1 X W T S V R U DMR15 D2 DMR16 DMR17 K DMR2 ...

Страница 32: ...pressure sensor during the inspiration phase The built in mechanical springs in the pressure sensor exert different forces on the diaphragms Non adjustable spring force 10 mbar Spring force adjustable 30 mbar to 60 mbar with the alarm control valve One or other of the diaphragms is raised according to the respective compen sation pressures on the mechanical springs Figure 19 Segment of the flow ch...

Страница 33: ...DMR7 10 Alarm control valve 19 Ventilator block 3 Alarm silence 11 Silence button 20 Frequency control 4 Alarm logic 12 Indicator Paw low 21 DMR1 DMR2 5 Volume frequency 13 Frequency control valve 22 Vent valve 6 Pressure sensor double diaphragm switch 14 Indicator Paw high 23 Pressure regulator 15 MV control valve 24 Patient connection port 7 DMR18 16 Indicator Pv 25 Gas connection port 8 Pressur...

Страница 34: ...For internal use only Copyright reserved Version 2 0_ Released_Printed on_15 03 06_F5503400T01 fm 30 5503 400 Function Description Oxylog 1000 ...

Страница 35: ...31 Maintenance Procedures ...

Страница 36: ...32 ...

Страница 37: ...rts according to hospital regulations prior to any servicing work or before returning the appliance or parts of it for repair pur poses NOTE Dispose of the pressure regulator in accordance with local waste disposal regulations Safety check Have routine safety checks performed every two years by trained specialists Inspectionandservicingof the appliance Have an inspection and service per formed eve...

Страница 38: ...connect the connector of the compressed gas hose from the socket on the central supply system compressed gas cylinder Detach the ventilation hose from the Oxylog 1000 Unscrew the compressed gas hose from the Oxylog 1000 Remove the scews 1 and washers Remve the grub screws 2 Withdraw the connection socket Fig 1 View of the Oxylog 1000 Removing screws Withdraw the drawer unit from the housing of the...

Страница 39: ...edures Remove the screws 1 Fig 2 Bottom view of the Oxylog 1000 Removing screws Detach the small silicone hose hose connection supply block to pres sure gauge from the hose connection socket of the supply block Plug the large silicone hose hose connection supply block to working block from the hose connection socket of the supply block 1 ...

Страница 40: ...5503 400 Maintenance Procedures Oxylog 1000 Press in the circlip on the hose connection socket at the same time with drawing the PA tube hose connection supply block to working block Figure 3 View of the drawer unit of the Oxylog 1000 Working block Pressure regulator Pressure gauge Supply block Circlip ...

Страница 41: ...hat the PA tube is seated firmly in the hose connection socket of the working block by carefully pulling on the PA tube Plug the large silicone hose hose connection supply block to working block onto the hose connection socket of the supply block Detach the small silicone hose hose connection supply block to pres sure gauge onto the hose connection socket of the supply block Mount the supply block...

Страница 42: ...nto the wall outlet com pressed gas cylinder Connect the ventilation hose with flowmeter measuring range 10 to 120 L min to the connection socket ventilation hose Figure 5 Oxylog 1000 with test set up Testing the supply pressure Set the frequency selector Freq to 10 1 min Set the minute volume adjuster MV to full right Set the ventilation pressure Paw to approx 60 mbar hPa Set the Air Mix No Air M...

Страница 43: ...wnstream pressure of the pressure regulator or counter clockwise to reduce the pressure After setting the correct downstream pressure press the adjusting ring of the pressure regulator down to prevent the downstream pressure being unintentionally altered Set the main switch of the Oxylog 1000 to O Detach the hose from the pressure gauge and re connect it to the con nection socket of the main switc...

Страница 44: ...For internal use only Copyright reserved Version 2 0_ Released_Printed on_15 03 06_K5503400V01 fm 40 5503 400 Maintenance Procedures Oxylog 1000 ...

Страница 45: ...41 Schematics and Diagrams ...

Страница 46: ...42 ...

Страница 47: ...cs and Diagrams 1 Flow chart of the Oxylog 1000 old Figure 1 Flow chart of the Oxylog 1000 old DMR1 DMR2 DMR3 DMR4 DMR5 DMR6 V1 V2 DR1 WV1 WV2 MV I NV SV DA D7 D6 DMR7 DMR8 DM DMR12 DMR13 DMR10 DMR9 DMR11 RV4 RV5 Z2 Z3 Z1 D3 BV PA S RV3 WV3 DMR18 EV S1 S2 A1 F1 DMR15 DMR16 DMR17 AG DMR14 V3 Z4 T1 D4 D2 D1 Insp Exp RV1 RV2 FD D5 VL ...

Страница 48: ...MR1 to DMR18 3 2 way valve DR Pressure regulator EV Vent valve F1 Filter FD Frequency control valve I Injector AG Alarm generator MV Minute volume control valve NV Emergency air valve PA Patient connection RV1 to RV5 Non return valve S1 ON OFF switch S2 Air Mix No Air Mix Selector S Hose system SV Safety valve T1 Pushbutton valve V1 to V3 Volume VL Manifold WV1 2 2 way valve changeover valve WV2 3...

Страница 49: ...he Oxylog 1000 new Figure 2 Flow chart of the Oxylog 1000 new FD V3 RV1 I WV1 WV2 EV SV MV D3 D4 RV2 T1 L1 V1 A1 F1 S1 DR1 VL VL1 S2 DM DMR18 NV WV3 PA BV S RV3 RV5 RV4 X W T S V R U DMR15 D2 DMR16 DMR17 DMR7 DMR8 DMR9 DMR10 DMR12 DMR13 DMR11 DMR14 F B G D1 C M L N H P K DMR1 DMR3 DMR2 Alarm logic Frequency control Silence control Ventilator block Pressure sensor ...

Страница 50: ... to DMR18 3 2 way valve DR Pressure regulator EV Vent valve F1 Filter FD Frequency control valve I Injector L1 Alarm generator MV Minute volume control valve NV Emergency air valve PA Patient connection RV1 to RV5 Non return valve S1 ON OFF switch S2 Air Mix No Air Mix Selector S Hose system SV Safety valve T1 Pushbutton valve V1 Volume VL Manifold WV1 2 2 way valve changeover valve WV2 3 2 way va...

Страница 51: ...ted on_15 03 06_S5503400T01 fm 47 Oxylog 1000 Schematics and Diagrams 3 Tubing Diagram of the Oxylog 1000 old Figure 3 Tubing diagram of the Oxylog 1000 old 10 11 12 13 9 8 6 7 5 4 3 1 2 P2 P1 V1 V2 V2 V1 FV Ex D2 D1 IN MS 1 MS 2 MS 3 MS 4 MS 6 MS 5 P M N K L I H G F E B D C A ...

Страница 52: ... Oxylog 1000 Table 3 Key to the tubing diagram of the Oxylog 1000 old 1 Silence control 2 ON OFF switch 3 Frequency control 4 Alarm logic 5 Patient block 6 Microswitch pin configuration 7 Volume 8 Pre tubing 9 Tubing for final assembly on total unit 10 Plug 11 Air Mix No Air Mix selector 12 Pushbutton valve silence 13 Operating block ...

Страница 53: ...agrams 4 Tubing Diagram of the Oxylog 1000 new Figure 4 Tubing diagram of the Oxylog 1000 new Hose 4x1 PAE nf 160 lg transp different signals red alarms yellow silence green insp brown exp blue Pcontrol switched orange Pcontrol not switched Hose 6x1 PE 100 lg Hose 2 7x0 65 PAE nf 200 lg Hose 7x2 5 Si nf 160 lg yellow bottom stenosis socket ...

Страница 54: ...For internal use only Copyright reserved Version 2 0_ Released_Printed on_15 03 06_S5503400T01 fm 50 5503 400 Schematics and Diagrams Oxylog 1000 ...

Страница 55: ...Annex Parts catalog Test List ...

Страница 56: ......

Страница 57: ...Parts catalog Oxylog 1000 Because you care Emergency Care Perioperative Care Critical Care Perinatal Care Home Care 5503 400 Revision 2006 03 14 09 56 56 Because you care ...

Страница 58: ...2 ...

Страница 59: ...log Item No Part No Description Qty Qty unit Remark 5703300 Caddy 1 000 St Revision 2006 03 14 09 56 56 5503 400 Items that are shown in the illustration but are not listed below the illustration are not available as spare parts 3 ...

Страница 60: ...5 Bag for CompactCaddy 1 000 St 4 5703306 PROTECTION COVER 1 000 St 5 5703307 Carrying Belt 1 000 St 6 5704216 All round Wall Holder 1 000 St 7 5704218 Wall Holder Adaptation Plate 1 000 St 8 5704217 Quick Power Connector 1 000 St Revision 2006 03 14 09 56 56 5503 400 Items that are shown in the illustration but are not listed below the illustration are not available as spare parts 4 ...

Страница 61: ...PROTECTION BAR LOW FOR CADDY 1 000 St 4 5703341 CYLINDER FIXATION SET 1 000 St 5 5703336 PROTECTION FOOT CADDY 1 000 St 6 5703340 GUIDING RAIL SET FOR CADDY 1 000 St 7 5703367 PROTECTION STRIP CADDY 1 000 St Revision 2006 03 14 09 56 56 5503 400 Items that are shown in the illustration but are not listed below the illustration are not available as spare parts 5 ...

Страница 62: ...3310 CLAW_FOR_CADDY 1 000 St 2 5703333 GUIDING RAIL FOR COMPACTCADDY 1 000 St 3 5703335 PROTECTION FOOT Compact CADDY 1 000 St Revision 2006 03 14 09 56 56 5503 400 Items that are shown in the illustration but are not listed below the illustration are not available as spare parts 6 ...

Страница 63: ...ck Power Connector 5704217 VALVE BOX 2M86785 Wall Holder Adaptation Plate 5704218 Caddy CLAW_FOR_CADDY 5703310 CYLINDER FIXATION SET 5703341 GUIDING RAIL SET FOR CADDY 5703340 PROTECTION BAR HIGH FOR CADDY 5703329 PROTECTION BAR LOW FOR CADDY 5703328 PROTECTION FOOT CADDY 5703336 PROTECTION STRIP CADDY 5703367 CompactCaddy CLAW_FOR_CADDY 5703310 GUIDING RAIL FOR COMPACTCADDY 5703333 PROTECTION FOO...

Страница 64: ...8 ...

Страница 65: ...Test List ...

Страница 66: ......

Страница 67: ...al condition of Oxylog 1000 5 2 Function tests 5 2 1 Testing supply pressure and safety valve 6 2 2 Testing vent valve 9 2 3 Testing I E ratio 10 2 4 Testing ventilatory frequency 10 2 5 Testing minute volume 11 2 6 Testing alarm logic and silence control 12 2 7 Testing O2 concentration 16 3 Deliver the Oxylog 1000 to the operators ready for operation 17 4 Tested 17 ...

Страница 68: ...ctional Tests 21 2 1 Checking the Supply Pressure and the Safety Valve 22 2 2 Checking the Vent Valve 25 2 3 Checking the I E Ratio 26 2 4 Checking the Ventilatory Frequency 26 2 5 Checking the Minute Volume 27 2 6 Checking Warning Logics and Silence Control 28 2 7 Checking the O2 Concentration 32 3 Supply Oxylog 1000 to user ready for operation 33 4 Tested 33 ...

Страница 69: ...tion blocks of Oxylog 1000 have been modified Therefore this Test List is divided in two parts Test List 1 Oxylog 1000 old and Test List 2 Oxylog 1000 new Oxylog 1000 new can be recognized by its serial number ARNL 0101 through 0120 and all units from serial number ARNN 0001 Before using the Test List check which type of unit you have and select the cor rect part of the Test List ...

Страница 70: ...For internal use only Copyright reserved Version 2 0_ Released_Printed on_15 03 06_K5503400TL0 fm 2 5503 400 Test List Oxylog 1000 ...

Страница 71: ...lace the inspections and maintenance work carried out by the manufacturer Serial no Device location 5503 400 Oxylog 1000 Test List Page 3 K5503400TL1 fm 15 03 06 Copyright reserved Reproduction permitted only for non commercial purposes Test List 1 Oxylog 1000 old Oxylog 1000 Folder no 5503 400 Issue 02 99 ...

Страница 72: ...15 03 06 Copyright reserved Reproduction permitted only for non commercial purposes Test list equipment Test equipment Measuring range Accuracy class Pressure gauge 30 to 120 mbar hPa 1 6 Pressure gauge 0 to 6 bar 1 6 Flowmeter 10 to 120 L min 1 6 Flowmeter 2 to 10 L min 1 6 Test pressure reducer Hoses div T piece Test lung O2 analyser e g Oxydig Adjustable resistor Oxygen measurement test adapter...

Страница 73: ...urposes 1 Checking general condition of Oxylog 1000 Check that the accompanying documents are with the Oxylog 1000 Check the following components for damage Housing of the Oxylog 1000 Ventilation pressure gauge Ventilation hose Ventilation valve Compressed gas connecting hose 1 5 m Compressed gas connecting hose 3 m 2 Function tests Carry out the following function tests with an O2 test pressure r...

Страница 74: ...ew the compressed gas connecting hose from the connection socket compressed gas supply Remove the drawer unit of the Oxylog 1000 by removing the screws with washers 1 Remove the screws 2 Withdraw the connection sockets ventilation hose and compressed gas supply Withdraw the drawer unit from the housing Fig 1 View of the Oxylog 1000 Removing the drawer unit Attach the connection socket compressed g...

Страница 75: ...nection socket of the main switch see following illustration Connect the ventilation hose with flowmeter measuring range 10 to 120 L min to the connection socket ventilation hose Fig 2 Oxylog 1000 with test set up Testing the supply pressure Set the frequency selector Freq to 10 1 min Set the minute volume adjuster MV to full right Set the ventilation pressure Paw to approx 60 mbar hPa Set the Air...

Страница 76: ... right Hold the connection socket ventilation hose closed with a plug or your thumb Set the main switch of the Oxylog 1000 to I The ventilation pressure of the Oxylog 1000 is less than or equal to 80 mbar hPa read off ventilation pressure from pressure gauge of Oxylog 1000 Set the main switch of the Oxylog 1000 to O Detach the compressed gas connector from the socket on the central supply compress...

Страница 77: ...ting vent valve Test set up Fig 4 Oxylog 1000 with test set up Testing the vent valve With the test pressure reducer feed a flow of 10 L min into the connection socket for the ventilation hose The maximum pressure rise is 8 mbar hPa read off the pressure rise from the pressure gauge of the Oxylog 1000 CAUTION To check the vent valve do not connect the compressed gas supply to the connection socket...

Страница 78: ...he maximum ventilation pressure Paw to full right Set the Air Mix No Air Mix selector switch to No Air Mix Switch on the Oxylog 1000 at the main switch Use a stopwatch to determine the I E ratio The I E ratio is 0 66 0 71 to 0 58 Formula to determine the I E ratio I E ratio 2 4 Testing ventilatory frequency Turn the maximum ventilation pressure Paw to full right Set the minute volume adjuster MV t...

Страница 79: ...ll right Set the frequency selector Freq to 10 1 min Set the minute volume adjuster MV variably see following table Set the required compensation pressure with the adjustable resistor during the inhalation phase Read off the compensation pressure from the pressure gauge of the Oxylog 1000 Determine the minute volume with and without enrichment of the ambient air see following table Minute volume a...

Страница 80: ...adjuster MV to approx 10 L min Set the frequency selector Freq to 10 1 min Set the ventilation pressure selector Paw to approx 40 mbar hPa Set the Air Mix No Air Mix selector switch to No Air Mix Slowly close the pressure regulator of the test pressure reducer Observe the pressure gauge of the test pressure reducer When the downstream pressure has reached a value of 2 2 bar to 1 9 bar the Pv indic...

Страница 81: ...y selector Freq to 10 1 min Set the Air Mix No Air Mix selector switch to Air Mix Set the ventilation pressure selector Paw to full right During the inhalation phase set a ventilation pressure of approx 30 mbar hPa with the adjustable resistor Note Read off the ventilation pressure from the pressure gauge of the Oxylog 1000 Slowly open the adjustable resistor during the inhalation phase until the ...

Страница 82: ... to 10 1 min Set the Air Mix No Air Mix selector switch to Air Mix Turn the ventilation pressure selector Paw to full right During the inhalation phase set a ventilation pressure of approx 30 mbar hPa with the adjustable resistor Note Read off the ventilation pressure from the pressure gauge of the Oxylog 1000 During the inhalation phase slowly close off the ventilation hose further with the adjus...

Страница 83: ...e muting Silence Test set up Fig 9 Oxylog 1000 with test set up Testing the muting Silence Set the minute volume adjuster MV to 10 L min Set the frequency selector Freq to 10 1 min Detach the test lung from the ventilation valve The Oxylog 1000 emits an audible alarm The Paw low indicator lights up red During the inhalation phase press the Alarm silence key The audible alarm stops The Alarm silenc...

Страница 84: ...he O2 concentration Set the frequency selector Freq to 10 1 min Set the ventilation pressure Paw to full right Set the Air Mix No Air Mix selector switch to Air Mix Set the minute volume adjuster MV to 7 L min The Oxydig shows 55 Vol O2 17 Vol O2 on the display Set the minute volume adjuster MV to 20 L min The Oxydig shows 55 Vol O2 17 Vol O2 on the display Set the Air Mix No Air Mix selector swit...

Страница 85: ...maintenance work carried out by the manufacturer 5503 400 Oxylog 1000 Test List Page 17 K5503400TL1 fm 15 03 06 Copyright reserved Reproduction permitted only for non commercial purposes 3 Deliver the Oxylog 1000 to the operators ready for operation 4 Tested Date _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ Name _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ ...

Страница 86: ...y available test aids and tools but does not replace the inspections and maintenance work carried out by the manufacturer 5503 400 Oxylog 1000 Test List Page 18 K5503400TL1 fm 15 03 06 Copyright reserved Reproduction permitted only for non commercial purposes ...

Страница 87: ... the inspections and maintenance work carried out by the manufacturer Serial no Installation site 5503 400 Oxylog 1000 Test List 2 Page 19 K5503400TL2 fm 15 03 06 Copyright reserved Reproduction permitted for non commercial purposes only Test List 2 Oxylog 1000 new Oxylog 1000 File no 5503 400 Edition 02 2000 ...

Страница 88: ... 03 06 Copyright reserved Reproduction permitted for non commercial purposes only Required Test Equipment Test equipment Range Accuracy class Pressure gauge 30 to 120 mbar hPa 1 6 Pressure gauge 0 to 6 bar 1 6 Flowmeter 10 to 120 L min 1 6 Flowmeter 2 to 10 L min 1 6 Test pressure regulator Hoses various T piece Test lung Oxygen analyzer e g Oxydig Adjustable restrictor Oxygen measurement test ada...

Страница 89: ...s only 1 Checking the General Condition of Oxylog 1000 Check that all accompanying documents to Oxylog 1000 are available Check the following components for possible damage Housing of Oxylog 1000 Ventilation pressure gauge Ventilation hose Ventilation valve Compressed air connecting hose 1 5 m Compressed air connecting hose 3 m 2 Functional Tests Carry out the following functional tests using an O...

Страница 90: ...on hose from the connecting piece ventilation hose Unscrew compressed gas connecting hose from connecting piece compressed gas supply To dismount the plug in unit of Oxylog 1000 remove the screws and washers 1 Remove the screws 2 Pull out the connecting piece ventilation hose and compressed gas supply Pull the plug in unit out of the housing Fig 1 Right front view of Oxylog 1000 Removing the plug ...

Страница 91: ... Connect ventilation hose with flowmeter measuring range 10 to 120 L min to connecting piece ventilation hose Fig 2 Oxylog 1000 with test set up Checking the supply pressure Set the frequency adjuster Freq to 10 1 min Set the minute volume adjuster MV to right limit stop Set the airway pressure Paw to right end stop Set the Air Mix No Air Mix selector to the No Air Mix position Place the ON OFF sw...

Страница 92: ... on the unit Fig 3 Checking the safety valve Reading on the pressure gauge Pplat 80 mbar r10 mbar Peak pressure Pmax 100 mbar Place the ON OFF switch of Oxylog 1000 to position O Remove the hose clamp Disconnect the compressed gas connector from the socket of the central gas supply system compressed gas cylinder Unscrew compressed gas connecting hose from connecting piece compressed gas supply Rem...

Страница 93: ...Checking the Vent Valve Test set up Fig 4 Oxylog 1000 with test set up Checking the vent valve Using the test pressure reducer feed a flow of 10 L min into the connecting piece of the ventilation hose The increase in pressure should not exceed 8 mbar hPa read pressure on pressure gauge of Oxylog 1000 CAUTION To check the vent valve do no connect the compressed gas supply to the connecting piece fo...

Страница 94: ...m airway pressure Paw to right end stop Set the Air Mix No Air Mix selector to the No Air Mix position Switch on Oxylog 1000 using the ON OFF switch Use the stopwatch to measure the I E ratio The I E ratio should be 0 625 to 0 833 Equation for determination of the I E ratio I E ratio 2 4 Checking the Ventilatory Frequency Set the maximum airway pressure Paw to right end stop Set the minute volume ...

Страница 95: ... frequency adjuster Freq to 10 1 min Set the minute volume adjuster MV to various steps see the following table Set the required counterpressure during the inspiratory phase using the adjustable restrictor Read counterpressure on pressure gauge of Oxylog 1000 Measure the minute volume with and without ambient air enrichment see the following table Minute volume adjuster Flow Counterpressure settin...

Страница 96: ...he minute volume adjuster MV to left limit stop Set the frequency adjuster Freq to 10 1 min Set the airway pressure adjuster Paw to approx 40 mbar hPa Set the Air Mix No Air Mix selector to the No Air Mix position Slowly close the pressure regulator of the test pressure reducer Watch the pressure gauge of the test pressure reducer As soon as the downstream pressure reaches 2 2 bar to 1 8 bar the P...

Страница 97: ...in Set the frequency adjuster Freq to 10 1 min Set the Air Mix No Air Mix selector to the Air Mix position Set the airway pressure adjuster Paw to right end stop During the inspiratory phase use the adjustable resistor to set an airway pressure of approx 30 mbar hPa Note Read airway pressure on pressure gauge of Oxylog 1000 During the inspiratory phase slowly open the adjustable resistor until the...

Страница 98: ...djuster Paw to right end stop During the inspiratory phase use the adjustable resistor to set an airway pressure of approx 30 mbar hPa Note Read airway pressure on pressure gauge of Oxylog 1000 During the inspiratory phase slowly close the ventilation hose further using the adjustable resistor until the indicator turns red Paw high The Paw high switching point is at 50 to 60 mbar hPa read the swit...

Страница 99: ...lence Test set up Fig 9 Oxylog 1000 with test set up Checking the alarm sound silence Set the minute volume adjuster MV to 10 L min Set the frequency adjuster Freq to 10 1 min Disconnect the test lung from the ventilation valve Oxylog 1000 should generate an audible alarm The Paw low indicator should turn red During the inspiratory phase press the silence button The audible alarm should stop The s...

Страница 100: ...ster Freq to 10 1 min Set the airway pressure Paw to right end stop Set the Air Mix No Air Mix selector to the Air Mix position Set the minute volume adjuster MV to 7 L min The oxygen analyzer Oxydig should show 55 vol O2 40 to 65 vol O2 on its display Set the minute volume adjuster MV to 20 L min The oxygen analyzer Oxydig should show 55 vol O2 40 to 65 vol O2 on its display Set the Air Mix No Ai...

Страница 101: ...s and maintenance work carried out by the manufacturer 5503 400 Oxylog 1000 Test List 2 Page 33 K5503400TL2 fm 15 03 06 Copyright reserved Reproduction permitted for non commercial purposes only 3 Supply Oxylog 1000 to user ready for operation 4 Tested Date _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ Name _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ ...

Страница 102: ... available test aids and tools but does not replace the inspections and maintenance work carried out by the manufacturer 5503 400 Oxylog 1000 Test List 2 Page 34 K5503400TL2 fm 15 03 06 Copyright reserved Reproduction permitted for non commercial purposes only ...

Страница 103: ......

Страница 104: ...Phone 49 0 1805 3723437 Fax 31 499 331 335 Fax 49 451 882 3779 Subject to change without notice Will not be replaced in the event of modifications Copyright by Dräger Medical AG Co KG Lübeck Germany The warranty and liability conditions of the general terms and conditions for business transactions of Dräger Medical AG Co KG are not extended by this Technical Documentation ...