This test list can be processed with standard commercially available test aids and tools, but does not replace the inspections and
maintenance work carried out by the manufacturer.
5503.400 Oxylog 1000
Test List
Page 6
K5503400TL1.fm
15.03.06
Co
pyri
ght
reser
ve
d
.
Repro
ducti
on p
e
rmitt
e
d o
n
ly f
o
r non-commerc
ial
purpo
ses.
2.1
Testing supply pressure and safety valve
• Set the main switch of the Oxylog 1000 to „0“.
• Detach the compressed gas connector from the socket on the central
supply/compressed gas cylinder.
• Unscrew the compressed gas connecting hose from the connection
socket (compressed gas supply).
• Remove the drawer unit of the Oxylog 1000 by removing the screws
with washers (1).
• Remove the screws (2).
• Withdraw the connection sockets (ventilation hose and compressed
gas supply).
• Withdraw the drawer unit from the housing.
Fig.
1:
View of the Oxylog 1000: Removing the drawer unit
• Attach the connection socket (compressed gas supply) by the screw to
the drawer unit of the Oxylog 1000.
• Attach the connection socket (ventilation hose) by the screw to the
drawer unit of the Oxylog 1000.
• Screw the compressed gas connecting hose onto the Oxylog 1000.
1
Connection socket
(ventilation hose)
2
Drawer unit
Connection socket
(compressed gas supply
Содержание Oxylog 1000
Страница 2: ......
Страница 5: ...1 General ...
Страница 6: ...2 ...
Страница 9: ...5 Function Description ...
Страница 10: ...6 ...
Страница 35: ...31 Maintenance Procedures ...
Страница 36: ...32 ...
Страница 45: ...41 Schematics and Diagrams ...
Страница 46: ...42 ...
Страница 55: ...Annex Parts catalog Test List ...
Страница 56: ......
Страница 58: ...2 ...
Страница 64: ...8 ...
Страница 65: ...Test List ...
Страница 66: ......
Страница 103: ......