2 • Description
Doc. 065215-02 9/09
9
performed. For example, an ICS-900 configured for anion analyses
uses carbonate eluent, while an ICS-900 configured for cation
analyses uses methanesulfonic acid (MSA) eluent.
•
Regenerant renews the suppressor’s ability to suppress eluent
conductivity. An ICS-900 configured for anion analyses uses dilute
sulfuric acid regenerant. An ICS-900 configured for cation analyses
uses tetrabutylammonium hydroxide (TBAOH) regenerant. For more
information about suppressor regeneration, see
2.1.2
Component Mounting Panel
shows the components installed on the component panel
behind the ICS-900 front door.
Figure 2-2. ICS-900 Component Mounting Panel
MMS 300
Suppressor
Injection Valve
Pressure
Transducer
Pump Heads
1
2
3
4
5
6
7
8
Separator Column
Guard Column
Tubing Chase
Conductivity Cell
(Housed in DS5
Detection Stabilizer
1
2
5
4
6
7
3
8
9
Backpressure
Tubing
9