3 • Operation and Maintenance
Doc. 065215-02 9/09
45
Autosampler Setup Notes
Follow the steps below to verify that the ICS-900 injection valve is
controlled by the correct device.
1. Open the Chromeleon Server Configuration program.
2. Open the ICS-900 Properties dialog box and click
Inject Valve
.
3. Under
Controlled by,
check the device selected for
Pump_InjectValve:
•
For an AS or AS-DV, verify that the
AS
•
For an AS40, verify that the
ICS-900
is shown (see
4. To change the controlling device, select the
Pump_InjectValve
name
and press the
F2
key. On the Device Configuration dialog box, select
the correct device from the
Controlled By
list. Click
OK
.
5. For an AS40, also verify that there is a relay connection from the
AS40 Load relay to the ICS-900
RELAY OUT 1
. Refer to the
for details.
Figure 3-12. ICS-900 Properties: Inject valve controlled by AS
Figure 3-13. ICS-900 Properties: Inject Valve Controlled by ICS-900
Содержание ICS-900
Страница 1: ...ICS 900 Ion Chromatography System Operator s Manual Document No 065215 Revision 02 September 2009...
Страница 120: ...ICS 900 Operator s Manual 112 Doc 065215 02 9 09...
Страница 124: ...ICS 900 Operator s Manual 116 Doc 065215 02 9 09...
Страница 132: ...ICS 900 Operator s Manual 124 Doc 065215 02 9 09...
Страница 142: ...ICS 900 Operator s Manual Index 10 Doc 065215 02 9 09...