ICS-900 Operator’s Manual
86
Doc. 065215-02 9/09
To complete the procedure:
1. Reconnect all liquid lines to the pump head.
2. Close the waste valve knob.
3. Turn on the ICS-900 power switch.
5. Reset the seal wear counter in Chromeleon or Chromeleon Xpress:
a. Press
F8
to open the Commands dialog box.
b. Expand the list of commands under
Pump_ECD
and scroll to the
ResetSealWear
c. Select the command and click
Execute
.
5.8
Replacing a Pump Piston
Continued leaking of the piston seal after installation of a new seal (assuming the
pump head is tight) indicates a dirty, scratched, or broken piston.
Follow the instructions in
to install a new piston (P/N 052840) and
piston seal (P/N 055870). Always replace the piston seal when replacing a piston.
After replacing the piston and seal, reset the seal wear and pump piston stroke
counters.