ICS-900 Operator’s Manual
68
Doc. 065215-02 9/09
Calibration
•
Pressure Transducer
•
Use the
Calibrate Offset
and
Calibrate Slope
buttons
when performing the pressure transducer calibration
procedure (see
•
The
Offset
and
Slope
values and the date of the last
calibration are displayed.
•
Conductivity Cell
•
Use the
Calibrate Offset
and
Calibrate Slope
buttons
when performing the cell calibration procedure (see
•
The
Offset
and
Slope
values and the date of the last
calibration are displayed.
•
Pump Flow Rate Calibration
•
Click
Calibration Procedure
to perform the calibration
•
Click
Log
to record the calibration value in the Audit
Trail.
Diagnostic
•
The
Pump
and
Conductivity Cell
Variance
values are a
measure of the noise of the respective channels. The values of
these measurements will vary with differing eluents, pump
seal age and wear, background conductivity, and so on. When
values (obtained under the same conditions) are compared
they can be used as a relative measure of the performance of
the system.
•
The
Variance
,
Minimum
, and
Maximum
readings for the
last time the
Pump
and
Conductivity
Cell
diagnostic tests
were run are displayed.
Dummy Cell
Test
•
The
Dummy Cell Test
connects a fixed resistance in place of
the conductivity cell. A measured value outside the typical
range of the test indicates a failure in the detector electronics.
•
Click
Log
to perform the test and record the value in the
Audit Trail. Clicking
Log
enables the dummy cell, logs the
value, and then disables the dummy cell.
•
The
Dummy Cell On
check box enables and disables the
dummy cell.
Содержание ICS-900
Страница 1: ...ICS 900 Ion Chromatography System Operator s Manual Document No 065215 Revision 02 September 2009...
Страница 120: ...ICS 900 Operator s Manual 112 Doc 065215 02 9 09...
Страница 124: ...ICS 900 Operator s Manual 116 Doc 065215 02 9 09...
Страница 132: ...ICS 900 Operator s Manual 124 Doc 065215 02 9 09...
Страница 142: ...ICS 900 Operator s Manual Index 10 Doc 065215 02 9 09...