ICS-900 Operator’s Manual
48
Doc. 065215-02 9/09
6. Use one of the following methods to inject the sample onto the
column:
•
Manually: click the
Inject
button on the ICS-900 Control panel.
See
for details about manually processing samples.
•
Automatically: Include the Inject command in a Chromeleon or
Chromeleon Xpress program. See the examples in
.
See
for details about automatically processing
samples.
3.11.6 Example Chromeleon Commands for Loading and
Injecting Samples
The following examples show commands for loading and injecting
samples using an AS, AS-DV, or AS40 autosampler.
Example AS Program Commands
Example AS-DV Program Commands
0.000
Pump_ECD.Autozero
;Zero the baseline.
Load
;Switch the valve to Load.
Wait
CycleTimeState ;Wait for cycle time (if any).
Inject
;Switch the valve to Inject.
Wait
InjectState
;Wait for injection to complete.
ECD_1.AcqOn
;Start data acquisition.
14.000
ECD_1.AcqOff
;Stop data acquisition.
Sampler.LoadPosition
;Switch the valve to Load.
DeliverSample
;Deliver sample to the valve.
0.000
Autozero
;Zero the baseline.
Wait
CycleTimeState ;Wait for cycle time (if any).
Inject
;Switch the valve to Inject.
ECD_1.AcqOn
;Start data acquisition.
14.000
ECD_1.AcqOff
;Stop data acquisition.
Содержание ICS-900
Страница 1: ...ICS 900 Ion Chromatography System Operator s Manual Document No 065215 Revision 02 September 2009...
Страница 120: ...ICS 900 Operator s Manual 112 Doc 065215 02 9 09...
Страница 124: ...ICS 900 Operator s Manual 116 Doc 065215 02 9 09...
Страница 132: ...ICS 900 Operator s Manual 124 Doc 065215 02 9 09...
Страница 142: ...ICS 900 Operator s Manual Index 10 Doc 065215 02 9 09...