4 • Troubleshooting
Doc. 065215-02 9/09
59
•
Liquid leaks at junction between pump head and pump housing
Replace the piston seal (see
4.4
Pump Does Not Start
•
No power (front door
Power
LED indicator fails to light)
Check that the power cord is plugged in.
Check the main power fuses and replace if needed (see
•
No communication between ICS-900 and Chromeleon or Chromeleon
Xpress
(
Link
LED on rear panel fails to light)
The USB cable is not connected correctly. For connection instructions, see the
ICS-900 Ion Chromatography System Installation Instructions
(Document
No. 065214), provided on the Dionex Reference Library CD-ROM (P/N
053891) and in the ICS-900 Ship Kit (P/N 067768).
4.5
No Flow
•
Pump not primed
•
Broken pump piston
Replace the piston (see
4.6
Erratic Flow/Pressure Reading
•
Pump needs priming
•
Damaged piston seal
Replace the piston seal (see
•
Dirty pump check valve
Clean or replace the check valve (see