5.11 Replacing the Conductivity Cell
1. Turn off the ICS-900 power switch.
1. Disconnect the power cord.
2. Open the ICS-900 front door.
3. Disconnect the black
ELUENT OUT
line from the
ELUENT OUT
port on the
4. Disconnect the orange
CELL OUT
line from the union that connects it to the
black
CELL OUT
line.
5. Use an Allen wrench to remove the two screws on the top of the cell housing.
6. Remove the cell by unplugging it from the component mounting panel.
7. Plug the new conductivity cell (P/N 067761) into its mounting location.
8. Replace the screws on the top of the cell housing and tighten.
9. Attach the orange
CELL OUT
line from the cell to the union attached to the
black
CELL OUT
line. Attach the
ELUENT OUT
line to the
ELUENT OUT
port on
the suppressor.
10. Close the front door.
Figure 5-15. Conductivity Cell Assembly
to suppressor
Connector to component
DS
5 D
ete
ctio
n S
tab
iliz
er
ELUENT OUT
mounting panel
CELL OUT (orange)
to CELL OUT (black)