ICS-900 Operator’s Manual
40
Doc. 065215-02 9/09
3.11.2 Manually Processing Samples
To manually process a sample, select operating parameters and
commands from the Chromeleon or Chromeleon Xpress panel tabset.
Commands are executed as soon as they are entered.
Summary of Manual Sample Processing
1. Complete the instructions in
prepare the ICS-900 for operation and to prepare the sample for
processing.
2. Load the sample into the injection valve sample loop with an
autosampler (see
3. On the ICS-900 Control panel, click the
Autozero
button.
4. On the
Sequence Control
panel, click the
Acq On
button (or click
the
Acquisition On/Off
button on the Chromeleon toolbar).
5. Switch the injection valve to the Inject position. The method used to
switch the injection valve depends on whether you are using an
autosampler (see
6. The signal plot is displayed on the ICS-900
Control panel (see
). Monitor the chromatogram and when sample data has
been collected, click the
Acq Off
button on the
Sequence Control