ICS-900 Operator’s Manual
62
Doc. 065215-02 9/09
•
Contaminated column
1. Clean the column as instructed in the column manual. Column manuals
are included on the Dionex Reference Library CD-ROM (P/N 053891).
2. If cleaning is unsuccessful, replace the column.
4.11 No Detector Response
•
Cell not properly installed
Check that the cell is plugged into the component mounting panel and the cell
housing is screwed down until the bottom of the housing is flush against the
sheet metal panel (see
and
•
No flow from pump
Several conditions may cause this condition; see
and
for details.
•
Cell electronics malfunctioning
Use the Wellness panel to test the electronics with a dummy cell (see
and the Chromeleon Help for instructions). If the conductivity
reading with the dummy cell is outside the tolerance range, the electronics are
malfunctioning. Contact Dionex for assistance.
4.12 High Detector Output
•
Background not suppressed by suppressor
If
Conductivity exceeds limit
is displayed in the Chromeleon or Chromeleon
Xpress Audit Trail, follow the instructions in
troubleshooting guidance, refer to the suppressor manual. Suppressor manuals
are included on the Dionex Reference Library CD-ROM (P/N 053891).
•
Sample concentration too high
Dilute the sample (see
•
Wrong eluent or regenerant
Check that you are using the correct eluent and regenerant for your system
(see
Содержание ICS-900
Страница 1: ...ICS 900 Ion Chromatography System Operator s Manual Document No 065215 Revision 02 September 2009...
Страница 120: ...ICS 900 Operator s Manual 112 Doc 065215 02 9 09...
Страница 124: ...ICS 900 Operator s Manual 116 Doc 065215 02 9 09...
Страница 132: ...ICS 900 Operator s Manual 124 Doc 065215 02 9 09...
Страница 142: ...ICS 900 Operator s Manual Index 10 Doc 065215 02 9 09...