ICS-900 Operator’s Manual
84
Doc. 065215-02 9/09
To install the new piston seal:
1. If this is the
secondary
pump head, open the waste valve knob by turning the
knob one-half turn counterclockwise.
2. Lubricate the seal and the pump head
opening with a small amount of isopropyl
alcohol to facilitate insertion.
3. Push the piston through the spacer and
then through the new piston seal.
4. Insert the piston and seal into the pump
head.
5. To seat the seal, push down on the spacer
until it is flush with the head. A clicking
sound indicates that the seal is correctly
seated.
Piston Seal
Spacer
Piston