ICS-900 Operator’s Manual
54
Doc. 065215-02 9/09
To troubleshoot:
1. If a sequence is being executed, terminate the sequence by selecting
Stop
on the
Batch
menu.
2. Turn off the ICS-900 power briefly and then restart.
3. Try to toggle the valve from Load to Inject by clicking the
Inject
button
on the ICS-900 Control panel.
4. If the problem persists, contact Dionex for assistance.
Various electronics-related problems can cause this error message to be
displayed.
If this warning appears, contact Dionex for assistance. The ICS-900
electronics components cannot be serviced by the user.
This error indicates a problem in the pump controller electronics.
To troubleshoot:
•
Contact Dionex for assistance. The ICS-900 electronics components
cannot be serviced by the user.
This error indicates a problem in the pump controller electronics.
To troubleshoot:
•
Contact Dionex for assistance. The ICS-900 electronics components
cannot be serviced by the user.
Module data buffer overflow. Data may have been lost.
Pump motor lost control.
Pump pressure hardware error.