ICS-900 Operator’s Manual
96
Doc. 065215-02 9/09
A.8 Column Heater/Thermostat
A.9 Conductivity Detector and Flow Cell
Void Volume
SRS 300 (4 mm): <50
μ
L
SRS 300 (2 mm): <15
μ
L
MMS 300 (4 mm): <50
μ
L
MMS 300 (2 mm): <15
μ
L
AMMS ICE 300 (4 mm): <50
μ
L
AMMS ICE 300 (2 mm): <15
μ
L
Anion AES: <35
μ
L
Cation AES: <35
μ
L
Sequential
Suppression for
Anions
With MMS and CRD 300
Suppressor Wear
Parts
None; peristaltic pump and inline filters not required
Column Heater/
Thermostat
Optional with CTS-10 external heater. Temperature set range of 25
to 85 ºC; minimum 5 ºC above ambient.
Type
Bipolar-heated conductivity cell; microprocessor-controlled digital
signal processing over a range of 0 to 10,000
μ
S/cm
Temperature
Compensation
Preset for accurate reading at 40 ºC
Cell Drive
8 kHz square wave
Auto Offset
-9,999 to 9,999
μ
S
Linearity
1%
Resolution
0.0047 nS/cm
Calibration
Factory-calibrated before shipment to the user; calibrated in the
field with memory retention and transfer