ICS-900 Operator’s Manual
114
Doc. 065215-02 9/09
D.4 How do I adjust retention times?
Retention times are calculated during calibration. The
Use Recently Detected
Retention Time
parameter in the Chromeleon QNT Editor (
General
tab) can be
used to compensate for some types of retention time drifts; for example,
evaporation of volatile components in pre-mixed solvents or an aging column.
Refer to the Chromeleon Help or user’s guide for details.
D.5 When should I remake standards?
Standards are used only for calibration and should always be made fresh (they
have a lifetime of only one week).
D.6 When should I remake eluents?
Eluents should be remade every two or three weeks. In addition, always remake
the regenerant and fill the regenerant bottle when you refill the eluent bottle.
D.7 How do I start Chromeleon or Chromeleon Xpress?
Click
Start
on the Windows taskbar, and then select
Programs > Chromeleon >
Chromeleon
.
D.8 How do I back up data?
In Chromeleon, select
File > Export/Backup
. Back up the data and indicate the
backup source. Chromeleon Xpress does not allow data to be saved.
D.9 How do I delete data?
In the Chromeleon Browser, highlight the sequence you want to delete and then
select
File > Delete
.
Содержание ICS-900
Страница 1: ...ICS 900 Ion Chromatography System Operator s Manual Document No 065215 Revision 02 September 2009...
Страница 120: ...ICS 900 Operator s Manual 112 Doc 065215 02 9 09...
Страница 124: ...ICS 900 Operator s Manual 116 Doc 065215 02 9 09...
Страница 132: ...ICS 900 Operator s Manual 124 Doc 065215 02 9 09...
Страница 142: ...ICS 900 Operator s Manual Index 10 Doc 065215 02 9 09...