ICS-900 Operator’s Manual
28
Doc. 065215-02 9/09
3.2
Turning On the ICS-900 Power
Press the power switch on the ICS-900 rear panel (see
system power. These are the conditions at power-up:
•
The pump is off.
•
The injection valve is in the Load position.
•
The conductivity detector begins reading the current conductivity.
3.3
Connecting to Chromeleon or Chromeleon Xpress
1. Turn on the PC.
2. Start the Chromeleon Server, if it is not already running.
a. Check the Chromeleon Server Monitor icon on the Windows taskbar.
•
When the server is running, the icon is gray
.
•
When the Server is not running, the icon is crossed out in red
. To
start the server, right-click the icon and select
Start Server
.
b. If the Server Monitor icon is not on the Windows taskbar, click
Start
and
select
All Programs
(or
Programs
)
> Chromeleon > Server Monitor
.
The Server Monitor opens. Click
Start
to start the server.
3. To start the Chromeleon client, click
Start
and select
All Programs
(or
Programs
)
> Chromeleon > Chromeleon
.
4. If Chromeleon is installed, display
the panel tabset by selecting
View >
Default Panel Tabset
or by
clicking the
Default Panel Tabset
toolbar button.
If Chromeleon Xpress is installed,
starting the application
automatically displays the panel tabset.
5. To display the ICS-900 Control panel, select the tab labeled with the ICS-900
timebase name (see
Содержание ICS-900
Страница 1: ...ICS 900 Ion Chromatography System Operator s Manual Document No 065215 Revision 02 September 2009...
Страница 120: ...ICS 900 Operator s Manual 112 Doc 065215 02 9 09...
Страница 124: ...ICS 900 Operator s Manual 116 Doc 065215 02 9 09...
Страница 132: ...ICS 900 Operator s Manual 124 Doc 065215 02 9 09...
Страница 142: ...ICS 900 Operator s Manual Index 10 Doc 065215 02 9 09...