E-DWT-H OPERATION AND MAINTENANCE MANUAL
© 2009 DH Instruments, a Fluke Company
Page 88
PRR
Purpose
Read the next available
Ready
condition, pressure measurement, rate and on-board
barometer reading.
Query “PRR?”
Classic “PRR”
Optional Suffix
“n”
A suffix may be used instead of “
RptLabel
”
‘1’ Specify the Hi Q-RPT
‘2’ Specify the Lo Q-RPT
Remarks The
next
available
Ready
condition, Q-RPT pressure measurement, rate of pressure change,
and barometric pressure is replied in the current pressure unit of measure. . This can take up
to the current read rate period (1.2 seconds by default). Each data field is separated by a
comma, and is returned in the following order:
ready, pressure UNITS, rate UNITS/s, atm UNITS
Here are the field descriptions:
ready:
Three character ready status field. Refer to the “SR” command.
pressure:
The measured pressure for the active Q-RPT in the current pressure unit.
This is followed by the current pressure unit.
rate:
The measured rate of pressure change for the active Q-RPT in the current
pressure unit per second. This is followed by the current pressure unit of measure.
atm:
The pressure measured by the RPM4-E-DWT on-board barometer in the current
pressure unit (and always absolute). This is followed by the current pressure
unit.
Example
(enhanced)
Query sent:
Query reply:
“PRR?”
“R,2306.265 kPag,0.011 kPa/s,97.000 kPa g”
Example
(classic)
Query sent:
Reply:
“PRR”
“R,2306.265 kPag,0.011 kPa/s,97.000 kPa g”
See Also
“PR”, “QPRR”, “SR”
3.6.1, 3.1.2, 3.9.7.1
QPRR
Purpose
Read the last Q-RPT pressure measurement, pressure rate and on-board barometer output
immediately.
Query “QPRR?”
Classic “QPRR”
Optional Suffix
“n”
A suffix may be used instead of “
RptLabel
”
‘1’ Specify the Hi Q-RPT
‘2’ Specify the Lo Q-RPT
Remarks
The last measured
Ready/Not Ready
condition, active Q-RPT pressure measurement, rate of
pressure change, and barometric pressure is replied in the current pressure unit of measure
immediately. This program message is useful when a rapid response of measured pressure
is needed. It does not result in quicker pressure measurements. Each data field is separated
by a comma, and is returned in the following order:
Ready, pressure UNITS, rate UNITS/s, atm UNITS
Here are the field descriptions:
ready:
Three character ready status field. Refer to the “SR” command.
pressure:
The measured pressure for the selected Q-RPT in the current pressure unit.
This is followed by the current pressure unit and measurement mode.
rate:
The measured rate of pressure change for the active Q-RPT in the current
unit of pressure per second. This is followed by the current pressure unit.
atm:
The pressure measured by the RPM4-E-DWT on-board barometer in the
current pressure unit (but always absolute). This is followed by the current
pressure unit.
Example
(enhanced)
Query sent:
Query reply:
“QPRR?”
“R,2306.265 kPa g,0.011 kPa/s,97.000 kPa g”
“R,2306.265 kPa g,0.011 kPa/s” (no barometer)
Example
(classic)
Query sent:
Reply:
“QPRR”
“R,2306.265 kPa g,0.011 kPa/s,97.000 kPa g”
See Also
“PR”, “PRR”, “SR”
3.6.1, 3.1.2, 3.9.7.1
Содержание E-DWT-H
Страница 6: ...E DWT H OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page IV Notes...
Страница 10: ...E DWT H OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page VIII Notes...
Страница 82: ...E DWT H OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 72 Notes...
Страница 120: ...E DWT H OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 110 Notes...
Страница 122: ...E DWT H OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 112 Notes...
Страница 124: ...E DWT H OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 114 Notes...