E-DWT-H OPERATION AND MAINTENANCE MANUAL
© 2009 DH Instruments, a Fluke Company
Page 46
•
Test point sequence: ascending, descending or both (Quick AutoTest only).
The initialization data is used to AutoRange the RPM4-E-DWT appropriately for the DUT and test
to be run. At the end of the AutoTest initialization, the RPM4-E-DWT is AutoRanged as follows
(see Section 3.7.5, Table 5):
•
RPT: The Lo Q-RPT is active if the maximum pressure in the AutoTest is equal to or less
than 1 000 psi (7 MPa), the Hi Q-RPT is active if the maximum pressure in the AutoTest
is greater than 1 000 psi (7 MPa).
•
Pressure unit of measure: As specified in the AutoTest initialization.
•
Display resolution: The DUT tolerance divided by 10 and rounded to the nearest digit, or
1 ppm of Q-RPT span, whichever is greater.
•
Stability test for
Ready/Not Ready
indication (see Section 3.1.2): The DUT tolerance
divided by 10, or 2 ppm of Q-RPT span, whichever is greater.
•
Upper limit (UL): The DUT full scale plus 5 %, or the Q-RPT’s maximum UL, whichever is
less (see Section 3.8.4).
To AutoRange the RPM4-E-DWT based on DUT characteristics but not execute an AutoTest, the
AutoTest can be aborted just after the initialization. Then the RPM4-E-DWT is set up
appropriately for the DUT to be tested but the AutoTest increments don’t execute.
See
OPERATION
in this Section for information on running a Quick or File AutoTest.
See Section 3.8.5.2 for information on setting up File AutoTest files.
See Section 3.8.5.1 for information on viewing the data files that result from running a File
AutoTest.
OPERATION
To run an AutoTest, press
[ENT]
from the Rate, RPT
or Clean run screen (DISPLAY modes that do not use
[ENT]
(see Section 3.7.7)). The display is:
Run AutoTest:
1Quick 2File
Select
<1Quick>
to run a Quick AutoTest or
<2File>
to run a File AutoTest (see
PRINCIPLE
above in this Section) and operation proceeds to test initialization.
3.7.11.1 AutoTest Initialization
When <2File> is selected to run a File AutoTest, the test initialization information
is obtained from the file selected. Operation proceeds directly to the test
confirmation. The RPM4-E-DWT asks the operator to confirm the range of the
AutoTest (<ENTER to run…>), AutoRanges (see Section 3.7.5) and operation
proceeds directly to test execution (see Section 3.7.11.2).
When
<1Quick>
is selected to run a
Quick AutoTest, the display is:
Run current Quick
AutoTest: 1yes 2no
Select
<1yes>
to rerun the last Quick AutoTest (see below in this Section).
Select
<2no>
to define a new Quick AutoTest.
When
<2no>
is selected from the
<Run current Quick AutoTest>
screen, the
AutoTest set-up/edit routine is accessed. The information needed to AutoRange
RPM4-E-DWT based on the DUT characteristics is then collected in successive
screens as follows.
Содержание E-DWT-H
Страница 6: ...E DWT H OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page IV Notes...
Страница 10: ...E DWT H OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page VIII Notes...
Страница 82: ...E DWT H OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 72 Notes...
Страница 120: ...E DWT H OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 110 Notes...
Страница 122: ...E DWT H OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 112 Notes...
Страница 124: ...E DWT H OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 114 Notes...