Security Levels
The security levels are structured to support typical operating environments as
follows:
None
This level is intended for use only by the system manager and/or
calibration facility. It allows access and editing in all areas including
critical metrological information.
Low
Low security is designed to protect the specific metrological
information and SYSTEM DIAGNOSTIC AND MAINTENANCE
functions of the system against accidental alteration. It is intended for an
advanced operator performing many different tasks. Low security is
the default user level setting.
Medium
Medium security is designed to protect specific metrological
information in the system and to assure that the RPM4-E-DWT is
operated using consistent operational parameters.
High
High security is designed to protect all operating parameters. It is
intended to minimize operator choices, for example to perform
repeated identical calibrations under consistent conditions.
Note
RPM4-E-DWT is delivered with the security level set at low to
avoid inadvertent altering of critical internal settings but with
access to changing security levels unrestricted. It is
recommended that the low security level be maintained at all
times. If there is a risk of unauthorized changing of the security
level, changing authority should be password protected (see
OPERATION of this section).
The security levels are structured to support typical levels of operation as shown
in Table 6. Specifically, the security levels prevent execution of the functions
accessed by the key strokes marked by “•”:
Table 6.
Security levels
FUNCTION
LOW
MEDIUM
HIGH
[RANGE]
•
[ENT] (AutoTest)
•
[UNIT]
•
[MODE]
•
[AutoRange]
• •
[LEAK CK]
•
[DISPLAY]
• •
[HEAD]
•
[SDS] (NOT USED IN THIS RPM4-E-DWT MODEL)
[AutoZ] (in absolute mode)
• •
[SETUP]
•
[SETUP], <1range>
• •
[SETUP], <2res>
• •
[SETUP], <4UL>
• •
[SETUP], <5ATest>
• •
[SPECIAL]
•
[SPECIAL], <1AutoZ>
• •
[SPECIAL], <1AutoZ>, <1on/1off>
• • •
[SPECIAL], <1AutoZ>, <3edit>
• • •
[SPECIAL], <2remote>
•
[SPECIAL], <2remote>,
make changes
• •
Содержание E-DWT-H
Страница 6: ...E DWT H OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page IV Notes...
Страница 10: ...E DWT H OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page VIII Notes...
Страница 82: ...E DWT H OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 72 Notes...
Страница 120: ...E DWT H OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 110 Notes...
Страница 122: ...E DWT H OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 112 Notes...
Страница 124: ...E DWT H OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 114 Notes...