3. OPERATION
Page 39
© 2009 DH Instruments, a Fluke Company
OPERATION
To activate the Rate DISPLAY press
[DISPLAY]
,
<2Rate>.
Pressing
<2Rate>
returns to the main run screen with the Rate DISPLAY active.
With the Rate DISPLAY active the main run screen is:
1. Current rate of change of pressure in current
pressure unit per second.
850.1 bar g Hi
R 0.1/sec
Note
•
The Rate DISPLAY is different and separate from the
stability setting which is used to set the stability
criterion on which the
Ready/Not Ready
indication is
based (see Sections 3.8.3 and 3.1.2). The Rate DISPLAY
only causes the current rate of change to be displayed
and has NO affect on the stability setting or the
Ready/Not Ready condition.
•
The Rate DISPLAY is the default RPM4-E-DWT main run
screen display.
•
To go to a DISPLAY other than Rate, press [DISPLAY]
and make a new DISPLAY choice (see Section 3.7.7).
3.7.7.3 DEV (Deviation)
PURPOSE
To activate the Deviation DISPLAY and/or set the deviation target value.
Note
See
Section
3.7.7,
Principle.
OPERATION
To activate the Deviation DISPLAY press
[DISPLAY]
,
<3Dev>
. The display is:
Target:
9000 psi g
Edit the desired target value. Pressing
[ENT]
returns to the main run screen with
the Deviation DISPLAY active using the entered target value.
Note
The target value is the value from which deviations (D) are
measured by the Deviation DISPLAY following:
D = current pressure - target pressure
With the Deviation DISPLAY active the main run screen is:
1. Deviation of current pressure from the target
value.
2. Target
value.
8998 psi g zH3
D -2 T9000.0
Содержание E-DWT-H
Страница 6: ...E DWT H OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page IV Notes...
Страница 10: ...E DWT H OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page VIII Notes...
Страница 82: ...E DWT H OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 72 Notes...
Страница 120: ...E DWT H OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 110 Notes...
Страница 122: ...E DWT H OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 112 Notes...
Страница 124: ...E DWT H OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 114 Notes...