4. REMOTE OPERATION
Page 79
© 2009 DH Instruments, a Fluke Company
4.4.4
Program Message Descriptions
#
Purpose
To allow the host PC to communicate with a device connected to the RPM4-E-DWT
COM2 port.
Classic
“#xx”
Arguments
xx:
The string to send out of the RPM4-E-DWT COM2 port. It must be less
than
40 characters long.
Remarks
The RPM4-E-DWT COM2 port can be used to communicate to another RS232 device
(such as another RPM4-E-DWT). This allows the user to use a single COM port or
IEEE-488 port on the host computer to communicate with the RPM4-E-DWT and another
RS232 device. A carriage return and a line feed (
<CR><LF>
) are added to the string.
After this program message is issued, the RPM4-E-DWT will reply back every string
received by the RPM4-E-DWT COM2 port that is terminated with a carriage return. Line
feeds are discarded. This will discontinue when the next query is sent to the
RPM4-E-DWT.
There is no other reply from this program message. Prior to using this program message,
you must ensure that the RPM4-E-DWT COM2 port is correctly set up to communicate
with the device on COM2. Refer to the
“COM2=“
program message.
Example
(classic)
Sent:
Reply:
“#VER”
“DH INSTRUMENTS RPM4-E-DWT us A70M/A7M Ver1.00 ”
This example assumes that a second RPM4-E-DWT’s COM1 port is connected to the
RPM4-E-DWT COM2 port. This example gets the version of the second RPM4-E-DWT.
See Also
“COM2”
3.9.2
ABORT
Purpose
Cancels any pending data requests by the “PR”, “PRR” “RATE”, “SR” or “CONT”
commands.
Command
Classic
“ABORT”
“ABORT”
Remarks
This program message is usually used to ensure the RPM4-E-DWT is in an idle condition.
Example
(enhanced)
Cmd sent:
Reply:
“ABORT”
“ABORT” (no reply if IEEE-488)
Example
(classic)
Sent:
Reply:
“ABORT”
“ABORT”
See Also
“PR”, “PRR”, “RATE”, “SR”, “CONT”
Содержание E-DWT-H
Страница 6: ...E DWT H OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page IV Notes...
Страница 10: ...E DWT H OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page VIII Notes...
Страница 82: ...E DWT H OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 72 Notes...
Страница 120: ...E DWT H OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 110 Notes...
Страница 122: ...E DWT H OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 112 Notes...
Страница 124: ...E DWT H OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 114 Notes...