E-DWT-H OPERATION AND MAINTENANCE MANUAL
© 2009 DH Instruments, a Fluke Company
Page 76
4.4
Commands
4.4.1
Programming Messages
Some commands accept an optional suffix. This suffix can be used to specify the explicit
Q-RPT to address. In most cases, each Q-RPT in the RPM4-E-DWT operates as an
independent measurement device. If the suffix is not specified, then the “Active” Q-RPT is
assumed.
The “RANGE” or “ARANGE” command can be used to select the “Active” Q-RPT. See the
program message descriptions for suffix use with a specific command.
Table 14.
Program message list
COMMAND
‘
n
’ indicates an
optional suffix
DESCRIPTION
#
Send a command string out of the RPM4-E-DWT COM2 port.
ABORT
Stop pending operations.
ARANGE
n
Read or set a new AutoRange to use.
ATESTDATAx,y
Read existing AutoTest data results.
ATESTRNGx
Read existing AutoTest range and hearder data.
ATM
Read the current atmospheric pressure (on-board barometer).
AUTOZERO
n
Read or set the status of the AutoZ automatic zeroing function.
AUTOZERO
n=RUN
Run AutoZero.
CALAMB
Read or set the on-board barometer calibration.
COM1
Read or set the configuration of the COM1 port.
COM2
Read or set the configuration of the COM2 port.
CONT
n
Enable continuous measurements.
DATE
Read or set the current date.
ERR
Read the last error message.
HEAD
Read or set the fluid head settings.
ID
Read or set the RPM4-E-DWT alphanumeric asset ID tag.
L2
Selects “classic” program message format.
L3
Selects “enhanced” program message format.
LOCAL
Return control to the RPM4-E-DWT front panel.
MEM
Read the power-up memory test status.
MSGFMT
Read or set the type of program message format to use.
PCAL
n
Read or set the user Lo Q-RPT calibration information.
PCAL:
XX
Read or set the user Lo Q-RPT calibration information (old command).
PR
n
Read the next RPM4-E-DWT pressure.
PRR
n
Read the next RPM4-E-DWT pressure, rate, and ATM.
QPRR
n
Read the last RPM4-E-DWT pressure, rate and ATM.
RANGE
Select a Q-RPT to be “active” in it’s default full scale range.
RATE
n
Read the next available rate of change of pressure.
READRATE
n
Read or set the Q-RPT measurement read rate and mode.
READYCK
n
Read or set a flag that is cleared by a
Not Ready
condition.
REMOTE
Enable remote local lockout operation.
RES
n
Read or set the pressure display resolution for the current Q-RPT and range.
RESET
Reset the RPM4-E-DWT to default user parameters.
RPT
n
Read the available Q-RPT data.
SCRSAV
Read or set the front panel screen saver period.
SN
Read the serial number of the RPM4-E-DWT.
SR
n
Read the next available pressure status (
Ready/Not Ready
).
SS
n
Read or set the stability required for a
Ready
condition.
SS%
n
Read or set the stability required for a
Ready
condition (% span/ s).
TIME
Read or set the current time of day.
UCOEF
n
Convert a pressure in Pascal to pressure in the current units.
UDU
Read or set the user defined pressure unit.
UL
n
Read or set the upper limit for the current range.
UNIT
n
Read or set the pressure unit of measure for the current range.
VER
Read the RPM4-E-DWT software version.
ZOFFSET
n
Read or set the AutoZero Z
offset
for the specified Q-RPT.
ZOFFSET:
XX
Read or set the AutoZ Z
offset
for the specified Q-RPT (old command).
Содержание E-DWT-H
Страница 6: ...E DWT H OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page IV Notes...
Страница 10: ...E DWT H OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page VIII Notes...
Страница 82: ...E DWT H OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 72 Notes...
Страница 120: ...E DWT H OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 110 Notes...
Страница 122: ...E DWT H OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 112 Notes...
Страница 124: ...E DWT H OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 114 Notes...