3. OPERATION
Page 21
© 2009 DH Instruments, a Fluke Company
A
Cooling fan
B
12VDC output connector (to E-DWT)
C
IEC power connection
D
ON/OFF switch (ON – battery connected to charger;
OFF - battery connected to 12 VDC output/E-DWT)
E
Charge level indicator
F
Charge level check switch
G
Battery charging indicator (green –
charging; green flashing – fully charged.
Figure 10.
Battery pack/charger
3.4
Selecting the Hi or Lo Q-RPT (Dual Q-RPT Models Only)
Note
This section applies only the E-DWT-H models with two Q-RPTs (see Section
1.2.3.1).
Selection of the active Q-RPT in the RPM4-E-DWT is made using the RPM4-E-DWT front panel’s
[RANGE]
key and E-DWT-H LO Q-RPT SHUT OFF valve.
When the Hi Q-RPT is active, the Lo Q-RPT must be isolated from the E-DWT’s pressure generation and
control system to avoid destroying it by overpressure.
The E-DWT’s Lo Q-RPT shut off valve and visual and audible indicators, assist the operator in performing
the correct operations to shut off and protect the Lo Q-RPT when the Hi Q-RPT is in use.
OPERATION
See Figure 12 for the schematic of E-DWT-H with dual Q-RPTs.
Numerical references in this Section refer to Figure 11.
The Lo Q-RPT is connected to or shut off from the rest of the E-DWT-H by the Lo Q-RPT shut off valve
that is operated using the valve knob. The proper valve position for the Q-RPT selected is prompted by
the
Valve Position LEDs
(7, 9).
The Lo Q-RPT shut off valve
(8) isolates and protects the Lo Q-RPT from the
TEST
port when it is
closed (knob fully CW) and opens the Lo Q-RPT to the
TEST
port when it is open (knob fully CCW).
Содержание E-DWT-H
Страница 6: ...E DWT H OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page IV Notes...
Страница 10: ...E DWT H OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page VIII Notes...
Страница 82: ...E DWT H OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 72 Notes...
Страница 120: ...E DWT H OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 110 Notes...
Страница 122: ...E DWT H OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 112 Notes...
Страница 124: ...E DWT H OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 114 Notes...