1. INTRODUCTION
Page 3
© 2009 DH Instruments, a Fluke Company
1.2.2
Battery and Charger Pack
The RPM4-E-DWT battery and charger pack is an optional accessory.
Power Requirements
100 to 240 VAC, 50/60 Hz, 15 W max consumption
Operating Temperature Range
10 to 40 °C
Storage Temperature Range
- 20 to 70 °C
Weight
2 kg (4.4 lb)
Dimensions
8 cm H x 22.5 cm W x 20 cm D (3.1 in. x 8.9 in. x 7.9 in.)
Battery Type
Nickel metal-hydride
Battery Voltage
12 VDC
Battery Capacity
Typical: 9000 mAh
Min:
8200
mAh
Charge Time
Full charge from empty, 14 to 16 hours approx.
Approx. Run Time, Full Charge
8 hours
1.2.3
Pressure Measurement Specifications
1.2.3.1 Reference Pressure Transducers (RPT)
The E-DWT’s RPM4-E-DWT is configured with one (Hi) or two (Hi and Lo) quartz
reference pressure transducer (Q-RPT) modules to measure pressure (see Table 1).
The Lo Q-RPT cannot be greater than A40M.
The Q-RPTs are intrinsically absolute, operating against a sealed vacuum
reference. Gauge pressure is defined by offsetting atmospheric pressure and
applying dynamic compensation for atmospheric changes using the on-board
barometer (see Section 3.9.1).
The complete E-DWT-H model definition includes the Q-RPTs included in its
RPM4-E-DWT. For example, an E-DWT-H with A200M and A20M Q-RPTs is
designated E-DWT-H A200M/A20M. An E-DWT-H with a single A200M Q-RPT
is designated E-DWT-H A200M.
Table 1.
Reference Pressure Transducer (Q-RPT) Module Designations and Ranges
SI
VERSION
US VERSION
Q-RPT
DESIGNATION
MAXIMUM
RANGE
[kPa]
Gauge
MAXIMUM
RANGE
[psi]
Gauge
A200M
1
200 000
30
000
A140M
1
140 000
20 000
A100M
1
100 000
15
000
A70M
1
70 000
10 000
A40M
1
40 000
6 000
A20M
1
20 000
3 000
A14M
1
14 000
2 000
A10M
1
10 000
1 500
A7M
1
7 000
1 000
Содержание E-DWT-H
Страница 6: ...E DWT H OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page IV Notes...
Страница 10: ...E DWT H OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page VIII Notes...
Страница 82: ...E DWT H OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 72 Notes...
Страница 120: ...E DWT H OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 110 Notes...
Страница 122: ...E DWT H OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 112 Notes...
Страница 124: ...E DWT H OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 114 Notes...