E-DWT-H OPERATION AND MAINTENANCE MANUAL
© 2009 DH Instruments, a Fluke Company
Page 86
LOCAL
Purpose
Returns control to the RPM4-E-DWT front panel.
Command “LOCAL”
Classic “LOCAL”
Remark
The REMOTE program message can lock the front panel out completely. The user can return
to local operation by sending the LOCAL program message, sending the IEEE-488 ‘GTL’
command (if in enhanced format), or by cycling RPM4-E-DWT power.
Example
(enhanced)
Cmd sent:
Reply:
“LOCAL”
“LOCAL”
(no reply if IEEE-488)
Example
(enhanced)
Cmd sent:
Reply:
“LOCAL?”
“LOCAL”
Example
(classic)
Sent:
Reply:
“LOCAL”
“LOCAL”
See Also
“REMOTE”
MEM
Purpose
Read the status from the power-up memory test.
Query “MEM?”
Classic “MEM”
Remarks
The RPM4-E-DWT system memory stores the user settings (units, resolution) and retains
them when the unit is OFF. On power-up, this memory is checked. If this memory is corrupted,
all user settings are reset to default , and the MEM status is set to reflect this.
Example
(enhanced)
Query sent:
Reply:
“MEM?”
“0” RPM4-E-DWT data corrupted and was set to factory defaults.
“1” The memory was found to be OK on power-up.
Example
(classic)
Sent:
Reply:
“MEM”
“MEM=0” RPM4-E-DWT data corrupted and was set to factory defaults.
“MEM=1” The memory was found to be OK on power-up.
See Also
3.9.9, 3.9.7.5
MSGFMT
Purpose
Read or set the type of program command format to use (enhanced or classic).
Command
Query
“MSGFMT
mode
“MSGFMT?”
Classic “MSGFMT=
mode
“MSGFMT”
Arguments
mode:
‘1’ to use the enhanced command format.
‘0’ to use the classic command format.
Defaults “MSGFMT
0”
Remarks
The user can select the type of remote command format to use. This format must agree with the
format sent to the RPM4-E-DWT
The enhanced query form of this command (“MSGFMT?
n
”) should always be used to set the
desired format, as it will be accepted regardless of the current format (classic or enhanced).
It is recommended to use the “L2” and “L3” commands instead of this command for new designs.
Example
(enhanced)
Cmd sent:
Query reply:
“MSGFMT 1”
“1”
(No reply if IEEE-488)
Example
(enhanced)
Cmd sent:
Query reply:
“MSGFMT? 1”
“1”
Example
(classic)
Sent:
Reply:
“MSGFMT=1”
“MSGFMT=1”
Errors
ERR# 6:
Missing or improper program message argument(s).
See Also
“L2 / L3”
4.3, 3.9.2.3
Содержание E-DWT-H
Страница 6: ...E DWT H OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page IV Notes...
Страница 10: ...E DWT H OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page VIII Notes...
Страница 82: ...E DWT H OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 72 Notes...
Страница 120: ...E DWT H OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 110 Notes...
Страница 122: ...E DWT H OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 112 Notes...
Страница 124: ...E DWT H OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 114 Notes...