E-DWT-H OPERATION AND MAINTENANCE MANUAL
© 2009 DH Instruments, a Fluke Company
Page 94
UDU
Purpose
Read or set the user defined pressure unit.
Command
Query
“UDU
label, ucoef
”
“UDU?”
Classic “UDU=
label, ucoef
”
“UDU”
label:
User unit label (4 alphanumeric char maximum). It cannot be an already
supported unit label.
Ucoef
:
“User unit conversion coefficient (units/Pa).
Default “UDU
USER,1.0”
Remarks
The user defined unit must be set up with the program message prior to remote or local
selection.
Example
(enhanced)
Cmd Sent:
Query reply:
“UDU MYUN, .001”
“MYUN, 0.001000”
(No reply if GPIB-488)
Example
(enhanced)
Cmd Sent:
Query reply:
“UDU? MYUN, .001”
“MYUN, 0.001000”
Example
(enhanced)
Sent:
Reply:
“UDU=MYUN, .001”
“MYUN, 0.001000”
Errors
See Also
3.9.6, 3.7.3
UL
n
Purpose
Read or set an upper limit for the specified or active Q-RPT.
Command
Query
“Ul
n
limit
”
“Ul
n
?”
Classic “Ul
n
=
limit
”
“UL
n
”
Optional Suffix
“n”
A suffix may be used instead of “
RptLabel
”
‘1’ Specify the Hi Q-RPT
‘2’ Specify the Lo Q-RPT
Arguments
“limit":
The upper limit pressure in the specified Q-RPT’s current pressure unit and
measurement mode.
Remarks
The RPM4-E-DWT has an upper limit for each range and for each measurement mode (gauge,
absolute, and differential). If the pressure does exceed the upper limit, the pressure display
flashes. This feature should always be used to prevent accidental over pressure of a device
under test.
Example
(enhanced)
Cmd sent:
Query reply:
“UL 1000”
“1000.00 kPa g”
Example
(enhanced)
Cmd sent:
Query reply:
“UL? 1000”
“1000.00 kPa g”
Example
(classic)
Sent:
Reply:
“UL=1000”
“1000.00 kPa g”
Errors
ERR# 6:
The “
limit
” argument is out of range.
See Also
3.8.4
Содержание E-DWT-H
Страница 6: ...E DWT H OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page IV Notes...
Страница 10: ...E DWT H OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page VIII Notes...
Страница 82: ...E DWT H OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 72 Notes...
Страница 120: ...E DWT H OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 110 Notes...
Страница 122: ...E DWT H OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 112 Notes...
Страница 124: ...E DWT H OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 114 Notes...