3. OPERATION
Page 43
© 2009 DH Instruments, a Fluke Company
PRINCIPLE
The RPM4-E-DWT measures gauge pressure at the height of the E-DWT’s top surface TEST
port. Frequently, when performing a calibration or test, the device or system under test is at
a different height than the E-DWT’s TEST port. This difference in height, frequently called
head
, can cause a significant difference between the pressure measured by the E-DWT-H at
its TEST port height and the pressure actually applied to the device under test at a different
height. When that is the case, it is useful to make a head correction to the pressure
measured by the RPM4-E-DWT in order to predict the pressure actually applied at a different
height.
RPM4-E-DWT can accurately determine “head” pressures for liquids (oil, water) as the
pressurized medium. In calculating the head value, standard gravity (9.80665 m/s
2
) is used.
A custom liquid density may also be specified.
The
[HEAD]
function key is used to specify the height difference between the E-DWT-H
TEST
port and another height. Entering a height of zero turns the function off. The height
units and the test medium can be changed using
[SPECIAL]
,
<3Head>
(see Section 3.9.3).
Note
An oil head has a pressure value of about 0.04 psi/inch (0.1 kPa/cm).
OPERATION
Note
For the head function to be used as described here, the RPM4-E-DWT
must be AutoZeroed to the TEST port reference level following the
AutoZeroing instructions in Section 3.7.10.
To access the HEAD function, press
[HEAD]
. The display is:
1.
Test fluid currently specified for the head correction.
2.
Edit field for head height.
Edit head height:
+ 14 in oil
Edit the head height to the desired value. Press
[ENT]
to return to the MAIN RUN screen
with the new head correction active. Press
[ESC]
to return with no changes.
Note
For the head function to be used as described
here, the RPM4-E-DWT must be AutoZeroed to
the TEST port reference level following the
AutoZeroing instructions in Section 3.7.10.
The reference height of the RPM4-E-DWT
pressure measurement is the E-DWT-H TEST
port weep hole. The head height should be
entered as a positive value if the device or
system under test is higher than the reference
level and negative if it is lower.
Содержание E-DWT-H
Страница 6: ...E DWT H OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page IV Notes...
Страница 10: ...E DWT H OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page VIII Notes...
Страница 82: ...E DWT H OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 72 Notes...
Страница 120: ...E DWT H OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 110 Notes...
Страница 122: ...E DWT H OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 112 Notes...
Страница 124: ...E DWT H OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 114 Notes...