5. MAINTENANCE, ADJUSTMENTS AND CALIBRATION
Page 105
© 2009 DH Instruments, a Fluke Company
5.4.4
RPM4-E-DWT Subassembly Description and Location
A
Micro
card
B
Driver
Board
C
On-Board Barometer
D
Lo Q-RPT module (optional)
E
Hi Q-RPT module
F
Display
Figure 15.
RPM4-E-DWT internal view
5.4.4.1 Micro Board
The micro board supports a Motorola 68302 micro-controller, EPROM,
EEPROM, 128k x 16 bit NVRAM, 8 Mbit flash memory; RS-232 and IEEE-488.2
communications; keypad and display control. An I/O port controls other ports
and devices in RPM4-E-DWT.
5.4.4.2 Driver Board
The driver board is controlled by the micro board (see Section 5.4.4.1). It
supports:
•
Frequency counters (2) for reading Q-RPTs (see Section 5.4.4.4)
•
On-board barometer power and output (see Section 5.4.4.3)
•
Power to the system cooling fan
•
Remote [ENT]
•
Keypad
and
display
•
Beeper