E-DWT-H OPERATION AND MAINTENANCE MANUAL
© 2009 DH Instruments, a Fluke Company
Page 30
3.6.4
Sounds
RPM4-E-DWT is equipped with a variable frequency tone device to provide audible feedback
and alarms. The beeper is used for the following indications.
Valid key press
Brief beep. Choice between three frequencies or NO
sound is available (see Section 3.9.5.2).
Invalid key press
Descending two tone “blurp”
Change E-DWT-H Lo Q-RPT
shut-off valve setting
Four one second beeps (see Section 3.4).
Leak check completed
Three two second beeps (see Section 3.7.6).
Upper limit exceeded
Intermittent one second beeps (see Section 3.8.4).
Pmax! (overpressure limit)
exceeded
Eight second high frequency beep (see Section 3.8.4.1).
AutoTest in/out of tolerance
reading
Ascending triad/descending triad (see Section 3.7.11).
Hi Q-RPT is selected but Lo
Q-RPT is being pressurized
(dual Q-RPT model RPM4-E-
DWT only)
Intermittent two second high frequency beep. The audible
alarm is combined with flashing of the red,
Lo Q-RPT
Active
LED on the E-DWT-H (see Section 3.4, 3.7.2).
AutoTest in/out of tolerance
reading
Ascending triad/descending triad (see Section 3.7.11).
3.7
RPM4-E-DWT Direct Function Keys
3.7.1
Summary
Note
Table 4 provides a brief summary of RPM4-E-DWT direct function key
operation. It may be useful to keep a copy of this summary near the
E-DWT-H, especially when first becoming acquainted with its operation.
Local operation of RPM4-E-DWT is through the front panel keypad. To minimize multi-
layered menu structures, the keypad numerical keys also provide direct access to the most
commonly used functions. The function accessed is labeled on the bottom half of the key.
Direct function keys are active whenever RPM4-E-DWT is in its MAIN RUN screen. Table 4
summarizes the operation of the direct function keys.
Содержание E-DWT-H
Страница 6: ...E DWT H OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page IV Notes...
Страница 10: ...E DWT H OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page VIII Notes...
Страница 82: ...E DWT H OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 72 Notes...
Страница 120: ...E DWT H OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 110 Notes...
Страница 122: ...E DWT H OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 112 Notes...
Страница 124: ...E DWT H OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 114 Notes...