AMPS-1 OPERATING MANUAL
11 Fetal ECG and Intrauterine
Pressure (optional)
11.1 Overview
AMPS-1 simulates fetal ECG and intrauterine pressures as observed during labour.
The simulated fetal signal represents a measurement from a fetal scalp electrode on the
normal ECG leads.
The intrauterine pressure (IUP) waveform simulates the measurements of a pressure
transducer during contractions. Transducer sensitivity can be set to either 5 or
40
ì
V/V/mmHg for this simulation.
11.2 Available Settings
Press ENTER after selecting a parameter to activate that setting.
11.2.1 Fetal Heart Rates
• 60, 90, 120, 150, 180, 210, and 240 BPM
11.2.2 Fetal Heart Rate Variation
• Uniform, early and late deceleration
11.2.3 Dynamic Intrauterine Pressure (IUP) waveform
• Positive bell-shaped pressure curve
11.2.4 Peak Intrauterine Pressure
• 25, 50 and 100 mmHg
11.2.5 Contraction Duration
• 90 seconds
Fetal/Intrauterine Pressure (Optional)/Chapter 11
#
Page 57
Содержание AMPS-1
Страница 1: ...AMPS 1 Advanc e d Mo dular Patie nt Sim ulato r Operating Manual ...
Страница 2: ......
Страница 4: ......
Страница 8: ......
Страница 14: ...AMPS 1 OPERATING MANUAL ...
Страница 44: ...AMPS 1 OPERATING MANUAL Arrhythmias Chapter 4 Page 30 ...
Страница 48: ...AMPS 1 OPERATING MANUAL Performance Waveforms Chapter 5 Page 34 ...
Страница 58: ...AMPS 1 OPERATING MANUAL Respiration Chapter 7 Page 44 ...
Страница 62: ...AMPS 1 OPERATING MANUAL Pacer Chapter 8 Page 48 ...
Страница 70: ...AMPS 1 OPERATING MANUAL Cardiac Output Optional Chapter 10 Page 56 ...
Страница 76: ...AMPS 1 OPERATING MANUAL Fetal Intrauterine Pressure Optional Chapter 11 Page 62 ...
Страница 80: ...AMPS 1 OPERATING MANUAL Figure 68 COM Port settings AMPSpc Companion Software Chapter 13 Page 66 ...
Страница 92: ...AMPS 1 OPERATING MANUAL Calibration and Maintenance Chapter 16 Page 78 ...
Страница 98: ...AMPS 1 OPERATING MANUAL Appendix A Page 84 ...
Страница 100: ......