AMPS-1 OPERATING MANUAL
BP1 RESP ARTIFACT
=<
ARTIFACT: OFF
Figure 41
- BP1 ADVANCED: ARTIFACT
Press the UP (
) or DOWN(
) arrow button to toggle the (power-on) default
sensitivity between 5 and 40 µV/V/mmHg.
Pressing the RIGHT (
u
) arrow button will change the menu screen to that of
Figure
41
.
Press the UP (
) or DOWN(
) arrow button to scroll through the settings provided
for the respiration artifact: OFF, 1-ON, and 2-ON.
Pressing the RIGHT (
u
) arrow button again will restore the LCD display to the initial
menu screen of
Figure 38
.
Press the LEFT (
t
) arrow button to scroll the display through the above simulator
parameters in reverse order.
Press ENTER after selecting a parameter to activate your selection.
NOTE: Dynamic BP waveforms may be incompatible with certain other functions, for
example, ECG performance waveforms. When incompatible ECG or other
simulations are selected, BP simulations will be set to static pressure at zero
mmHg on all channels.
Invasive Blood Pressure/Chapter 6
#
Page 40
Содержание AMPS-1
Страница 1: ...AMPS 1 Advanc e d Mo dular Patie nt Sim ulato r Operating Manual ...
Страница 2: ......
Страница 4: ......
Страница 8: ......
Страница 14: ...AMPS 1 OPERATING MANUAL ...
Страница 44: ...AMPS 1 OPERATING MANUAL Arrhythmias Chapter 4 Page 30 ...
Страница 48: ...AMPS 1 OPERATING MANUAL Performance Waveforms Chapter 5 Page 34 ...
Страница 58: ...AMPS 1 OPERATING MANUAL Respiration Chapter 7 Page 44 ...
Страница 62: ...AMPS 1 OPERATING MANUAL Pacer Chapter 8 Page 48 ...
Страница 70: ...AMPS 1 OPERATING MANUAL Cardiac Output Optional Chapter 10 Page 56 ...
Страница 76: ...AMPS 1 OPERATING MANUAL Fetal Intrauterine Pressure Optional Chapter 11 Page 62 ...
Страница 80: ...AMPS 1 OPERATING MANUAL Figure 68 COM Port settings AMPSpc Companion Software Chapter 13 Page 66 ...
Страница 92: ...AMPS 1 OPERATING MANUAL Calibration and Maintenance Chapter 16 Page 78 ...
Страница 98: ...AMPS 1 OPERATING MANUAL Appendix A Page 84 ...
Страница 100: ......