AMPS-1 OPERATING MANUAL
Figure 12
- QRS Complex
3.3 Available Settings
Press ENTER after selecting a parameter to activate that setting.
3.3.1 Rates
• 30, 40, 60, 80, 100, 120, 140, 160, 180, 200, 220, 240, 260, 280 and 300 BPM
3.3.2 Amplitudes (Lead II)
• 0.1, 0.25, 0.5, 1, 2, 3, 4 and 5 mV - using normal settings
• 0.05 to 5.50 mV in 0.05 mV steps - using performance settings
3.3.3 Baseline (Electrode) Impedance
• 500, 1000, 1500 and 2000
Ù
.
3.3.4 Artifact selections
• 50Hz, 60Hz, Muscle, Wandering and OFF. "Wandering" artifact uses the
respiration simulation settings for rate and waveshape.
3.3.5 ST Segments (Lead II)
• 0 mV (normal)
• Elevated, +0.1 to +0.8mV in +0.1 mV steps
• Depressed, -0.1 to -0.8mV in -0.1 mV steps
3.3.6 Axis Deviation
• Normal (intermediate), horizontal, and vertical. Modifies ECG baseline during
arrhythmias.
3.3.7 Neonatal Mode
• ECG R-wave width is reduced to 40 msec
ECG Normal Sinus Rhythm/Chapter 3
#
Page 16
Содержание AMPS-1
Страница 1: ...AMPS 1 Advanc e d Mo dular Patie nt Sim ulato r Operating Manual ...
Страница 2: ......
Страница 4: ......
Страница 8: ......
Страница 14: ...AMPS 1 OPERATING MANUAL ...
Страница 44: ...AMPS 1 OPERATING MANUAL Arrhythmias Chapter 4 Page 30 ...
Страница 48: ...AMPS 1 OPERATING MANUAL Performance Waveforms Chapter 5 Page 34 ...
Страница 58: ...AMPS 1 OPERATING MANUAL Respiration Chapter 7 Page 44 ...
Страница 62: ...AMPS 1 OPERATING MANUAL Pacer Chapter 8 Page 48 ...
Страница 70: ...AMPS 1 OPERATING MANUAL Cardiac Output Optional Chapter 10 Page 56 ...
Страница 76: ...AMPS 1 OPERATING MANUAL Fetal Intrauterine Pressure Optional Chapter 11 Page 62 ...
Страница 80: ...AMPS 1 OPERATING MANUAL Figure 68 COM Port settings AMPSpc Companion Software Chapter 13 Page 66 ...
Страница 92: ...AMPS 1 OPERATING MANUAL Calibration and Maintenance Chapter 16 Page 78 ...
Страница 98: ...AMPS 1 OPERATING MANUAL Appendix A Page 84 ...
Страница 100: ......